Cargando…
The HARP resistive plate chambers: Characteristics and physics performance
The HARP resistive plate chamber (RPC) system was designed for time-of-flight measurement in the large-angle acceptance region of the HARP spectrometer. It comprised 46 four-gap glass RPCs covering an area of . The design of the RPCs, their operation, intrinsic properties, and system performance are...
Autores principales: | , , , , , , , , , , , , , , , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2009
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1016/j.nima.2008.12.213 http://cds.cern.ch/record/1280438 |
Sumario: | The HARP resistive plate chamber (RPC) system was designed for time-of-flight measurement in the large-angle acceptance region of the HARP spectrometer. It comprised 46 four-gap glass RPCs covering an area of . The design of the RPCs, their operation, intrinsic properties, and system performance are described. The intrinsic time resolution of the RPCs is 127 ps leading to a system time-of-flight resolution of 175 ps. It is argued that gross differences between pion and proton timing in multi-gap timing RPCs, reported recently in the literature, do not exist. |
---|