Cargando…
Proportional counters aged anode wire recovering using an 80%CF4 + 20%CO2 gas mixture
A technique to recover a gas proportional counter having an aged anode wire using a glow discharge in an 80%CF4 + 20%CO2 gas mixture has been developed and tested. Studies of aging effects were carried out under sustained irradiation by an intense 90Sr -source of the straw proportional counters op...
Autores principales: | , , , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2011
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/1353259 |
Sumario: | A technique to recover a gas proportional counter having an aged anode wire using a glow discharge in an 80%CF4 + 20%CO2 gas mixture has been developed and tested. Studies of aging effects were carried out under sustained irradiation by an intense 90Sr -source of the straw proportional counters operated with a 60%Ar + 30%CO2 + 10%CF4 gas mixture. Special attention was paid to the aging mechanism of the anode wires. Our experience showed that using a given gas mixture the swelling of the anode wires is a typical mode of aging that leads to degradation of the gas gain. The proposed method of recovery provided a complete restoration of the gas gain and the signal amplitude in the damaged zone of the wire. SEM/XEM analysis confirmed successful cleaning WOx deposits from the wire surface. The application of this method to recover the aged gaseous detectors in real experimental conditions is discussed. |
---|