Cargando…

A new neutron monitor with silver activation

A moderator-type neutron monitor has been developed, which registers delayed beta rays from neutron-induced silver activation and which is able to measure dose equivalent in pulsed fields with peak dose rates of several thousand Sv h(-1). The monitor uses four silicon diodes in the centre of a polye...

Descripción completa

Detalles Bibliográficos
Autores principales: Luszik-Bhadra, M, Otto, Th, Hohmann, E
Lenguaje:eng
Publicado: 2010
Materias:
Acceso en línea:https://dx.doi.org/10.1016/j.radmeas.2010.06.009
http://cds.cern.ch/record/1359264
Descripción
Sumario:A moderator-type neutron monitor has been developed, which registers delayed beta rays from neutron-induced silver activation and which is able to measure dose equivalent in pulsed fields with peak dose rates of several thousand Sv h(-1). The monitor uses four silicon diodes in the centre of a polyethylene moderator, 30 cm in diameter. Two of the diodes are covered by natural silver foils and two of them by tin foils. The latter are used to subtract photon-induced pulses. For registering signals, a pulse height threshold is set at 662 key, which minimizes the effect of Cs-137 and lower energy radiation and - in addition - enhances the detection of beta rays from the shorter half-life silver isotope Ag-110 (25 s) as compared to the longer half-life isotope Ag-108 (144 s). The results of measurements in neutron and photon calibration fields, of MCNPX neutron response calculations and of first measurements in a high-intensity pulsed field at the PSI accelerator are shown. (c) 2010 Elsevier Ltd. All rights reserved.