Cargando…

Development of a Silicon Detector Monitor for the Superconducting Upgrade of the REX-ISOLDE Heavy-Ion Linac at CERN

A silicon detector monitor has been developed and tested in the frame of the beam diagnostics development program for the HIE-ISOLDE superconducting upgrade of the REX-ISOLDE heavy-ion linac at CERN. The monitor is intended for beam energy and timing measurements as well as for phase scanning of the...

Descripción completa

Detalles Bibliográficos
Autores principales: Zocca, F, Bravin, E, Voulot, D, Wenander, F, Pasini, M, Fraser, MA
Lenguaje:eng
Publicado: 2011
Materias:
Acceso en línea:http://cds.cern.ch/record/1372165
Descripción
Sumario:A silicon detector monitor has been developed and tested in the frame of the beam diagnostics development program for the HIE-ISOLDE superconducting upgrade of the REX-ISOLDE heavy-ion linac at CERN. The monitor is intended for beam energy and timing measurements as well as for phase scanning of the superconducting cavities. Tests have been performed with a stable ion beam, composed of carbon, oxygen and neon ions accelerated to energies from 300 keV/u to 2.85 MeV/u. The energy measurements performed allowed for beam spectroscopy and ion identification with a resolution of 1.3 % rms. The achieved resolution is suited for cavity phase scanning, which was demonstrated with the REX 7-gap resonator. The time structure of the beam, characterized by a bunch period of 9.87 ns, was measured with a resolution better than 200 ps. This paper describes the results from all these tests as well as providing details of the detector setup.