CALIBRATION ERRORS IN THE CAVITY BEAM POSITION MONITOR SYSTEM AT THE ATF2
It has been shown at the Accelerator Test Facility at KEK, that it is possible to run a system of 37 cavity beam position monitors (BPMs) and achieve high working resolution. However, stability of the calibration constants (position scale and radio frequency (RF) phase) over a three/four week runnin...
Autores principales: | , , , |
---|---|
Formato: | info:eu-repo/semantics/article |
Lenguaje: | eng |
Publicado: |
2011
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/1403143 |
_version_ | 1780923718607831040 |
---|---|
author | Cullinan, F Boogert, S T Joshi, N Lyapin, A |
author_facet | Cullinan, F Boogert, S T Joshi, N Lyapin, A |
author_sort | Cullinan, F |
collection | CERN |
description | It has been shown at the Accelerator Test Facility at KEK, that it is possible to run a system of 37 cavity beam position monitors (BPMs) and achieve high working resolution. However, stability of the calibration constants (position scale and radio frequency (RF) phase) over a three/four week running period is yet to be demonstrated. During the calibration procedure, random beam jitter gives rise to a statistical error in the position scale and slow orbit drift in position and tilt causes systematic errors in both the position scale and RF phase. These errors are dominant and have been evaluated for each BPM. The results are compared with the errors expected after a tested method of beam jitter subtraction has been applied. |
format | info:eu-repo/semantics/article |
id | cern-1403143 |
institution | Organización Europea para la Investigación Nuclear |
language | eng |
publishDate | 2011 |
record_format | invenio |
spelling | cern-14031432019-09-30T06:29:59Z http://cds.cern.ch/record/1403143 eng Cullinan, F Boogert, S T Joshi, N Lyapin, A CALIBRATION ERRORS IN THE CAVITY BEAM POSITION MONITOR SYSTEM AT THE ATF2 Accelerators and Storage Rings 9: Technology for normal conducting higher energy linear accelerators It has been shown at the Accelerator Test Facility at KEK, that it is possible to run a system of 37 cavity beam position monitors (BPMs) and achieve high working resolution. However, stability of the calibration constants (position scale and radio frequency (RF) phase) over a three/four week running period is yet to be demonstrated. During the calibration procedure, random beam jitter gives rise to a statistical error in the position scale and slow orbit drift in position and tilt causes systematic errors in both the position scale and RF phase. These errors are dominant and have been evaluated for each BPM. The results are compared with the errors expected after a tested method of beam jitter subtraction has been applied. info:eu-repo/grantAgreement/EC/FP7/227579 info:eu-repo/semantics/openAccess Education Level info:eu-repo/semantics/article http://cds.cern.ch/record/1403143 2011 |
spellingShingle | Accelerators and Storage Rings 9: Technology for normal conducting higher energy linear accelerators Cullinan, F Boogert, S T Joshi, N Lyapin, A CALIBRATION ERRORS IN THE CAVITY BEAM POSITION MONITOR SYSTEM AT THE ATF2 |
title | CALIBRATION ERRORS IN THE CAVITY BEAM POSITION MONITOR SYSTEM AT THE ATF2 |
title_full | CALIBRATION ERRORS IN THE CAVITY BEAM POSITION MONITOR SYSTEM AT THE ATF2 |
title_fullStr | CALIBRATION ERRORS IN THE CAVITY BEAM POSITION MONITOR SYSTEM AT THE ATF2 |
title_full_unstemmed | CALIBRATION ERRORS IN THE CAVITY BEAM POSITION MONITOR SYSTEM AT THE ATF2 |
title_short | CALIBRATION ERRORS IN THE CAVITY BEAM POSITION MONITOR SYSTEM AT THE ATF2 |
title_sort | calibration errors in the cavity beam position monitor system at the atf2 |
topic | Accelerators and Storage Rings 9: Technology for normal conducting higher energy linear accelerators |
url | http://cds.cern.ch/record/1403143 http://cds.cern.ch/record/1403143 |
work_keys_str_mv | AT cullinanf calibrationerrorsinthecavitybeampositionmonitorsystemattheatf2 AT boogertst calibrationerrorsinthecavitybeampositionmonitorsystemattheatf2 AT joshin calibrationerrorsinthecavitybeampositionmonitorsystemattheatf2 AT lyapina calibrationerrorsinthecavitybeampositionmonitorsystemattheatf2 |