Response of CVD Diamond Detectors to 14 MeV Neutrons

A series of measurements was taken at the Physikalisch-Technische Bundesanstalt (PTB) Braunschweig [1] using the 14 MeV neutron beam at the Van der Graaf accelerator with chemical vapor deposition (CVD) diamond detectors, in preparation of an upcoming (n, ) cross-section measurement [2] at the CERN-...

Descripción completa

Detalles Bibliográficos
Autores principales: Weiss, C, Griesmayer, E, Gagnon-Moisan, F, Kasper, A, Lucke, A, Schuhmacher, H, Weierganz, M, Zimba, A
Lenguaje:eng
Publicado: 2012
Materias:
Acceso en línea:http://cds.cern.ch/record/1494167
Descripción
Sumario:A series of measurements was taken at the Physikalisch-Technische Bundesanstalt (PTB) Braunschweig [1] using the 14 MeV neutron beam at the Van der Graaf accelerator with chemical vapor deposition (CVD) diamond detectors, in preparation of an upcoming (n, ) cross-section measurement [2] at the CERN-n TOF experiment [3, 4]. A single-crystal (sCVD) as well as a poly-crystalline (pCVD) diamond detector were used for the measurements. The response of both materials to the mono-energetic neutron beam was studied, also with the prospect for future applications in plasma diagnostics for fusion research. The results of the measurements are presented in this report.