Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor

The cernox™ sensor is delivered with a 3-point resistance versus temperature cal-ibration that permits the construction of an individual interpolation table by using the data in the CERN thermometer database. For instance at the 4.2 K point, the individual calibration and the manufacturer data are w...

Descripción completa

Detalles Bibliográficos
Autores principales: Balle, C, Casas, J, Fortescue-Beck, E, Vauthier, N
Lenguaje:eng
Publicado: 2013
Materias:
Acceso en línea:http://cds.cern.ch/record/1507610
_version_ 1780927524539203584
author Balle, C
Casas, J
Fortescue-Beck, E
Vauthier, N
author_facet Balle, C
Casas, J
Fortescue-Beck, E
Vauthier, N
author_sort Balle, C
collection CERN
description The cernox™ sensor is delivered with a 3-point resistance versus temperature cal-ibration that permits the construction of an individual interpolation table by using the data in the CERN thermometer database. For instance at the 4.2 K point, the individual calibration and the manufacturer data are within +/-0.1 K for 99.39% of a sample population of about 5700 sensors. Preliminary results also indicate that accuracies of 0.1 K and 1 K can be obtained below respectively 5 K and 77 K.
id cern-1507610
institution Organización Europea para la Investigación Nuclear
language eng
publishDate 2013
record_format invenio
spelling cern-15076102019-09-30T06:29:59Zhttp://cds.cern.ch/record/1507610engBalle, CCasas, JFortescue-Beck, EVauthier, NAccuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ SensorAccelerators and Storage RingsThe cernox™ sensor is delivered with a 3-point resistance versus temperature cal-ibration that permits the construction of an individual interpolation table by using the data in the CERN thermometer database. For instance at the 4.2 K point, the individual calibration and the manufacturer data are within +/-0.1 K for 99.39% of a sample population of about 5700 sensors. Preliminary results also indicate that accuracies of 0.1 K and 1 K can be obtained below respectively 5 K and 77 K.CERN-ATS-2013-003oai:cds.cern.ch:15076102013-01-22
spellingShingle Accelerators and Storage Rings
Balle, C
Casas, J
Fortescue-Beck, E
Vauthier, N
Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor
title Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor
title_full Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor
title_fullStr Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor
title_full_unstemmed Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor
title_short Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor
title_sort accuracy of the approximation function deduced from the fixed 3-points calibration delivered with the cernox™ sensor
topic Accelerators and Storage Rings
url http://cds.cern.ch/record/1507610
work_keys_str_mv AT ballec accuracyoftheapproximationfunctiondeducedfromthefixed3pointscalibrationdeliveredwiththecernoxsensor
AT casasj accuracyoftheapproximationfunctiondeducedfromthefixed3pointscalibrationdeliveredwiththecernoxsensor
AT fortescuebecke accuracyoftheapproximationfunctiondeducedfromthefixed3pointscalibrationdeliveredwiththecernoxsensor
AT vauthiern accuracyoftheapproximationfunctiondeducedfromthefixed3pointscalibrationdeliveredwiththecernoxsensor