Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor
The cernox™ sensor is delivered with a 3-point resistance versus temperature cal-ibration that permits the construction of an individual interpolation table by using the data in the CERN thermometer database. For instance at the 4.2 K point, the individual calibration and the manufacturer data are w...
Autores principales: | , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2013
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/1507610 |
_version_ | 1780927524539203584 |
---|---|
author | Balle, C Casas, J Fortescue-Beck, E Vauthier, N |
author_facet | Balle, C Casas, J Fortescue-Beck, E Vauthier, N |
author_sort | Balle, C |
collection | CERN |
description | The cernox™ sensor is delivered with a 3-point resistance versus temperature cal-ibration that permits the construction of an individual interpolation table by using the data in the CERN thermometer database. For instance at the 4.2 K point, the individual calibration and the manufacturer data are within +/-0.1 K for 99.39% of a sample population of about 5700 sensors. Preliminary results also indicate that accuracies of 0.1 K and 1 K can be obtained below respectively 5 K and 77 K. |
id | cern-1507610 |
institution | Organización Europea para la Investigación Nuclear |
language | eng |
publishDate | 2013 |
record_format | invenio |
spelling | cern-15076102019-09-30T06:29:59Zhttp://cds.cern.ch/record/1507610engBalle, CCasas, JFortescue-Beck, EVauthier, NAccuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ SensorAccelerators and Storage RingsThe cernox™ sensor is delivered with a 3-point resistance versus temperature cal-ibration that permits the construction of an individual interpolation table by using the data in the CERN thermometer database. For instance at the 4.2 K point, the individual calibration and the manufacturer data are within +/-0.1 K for 99.39% of a sample population of about 5700 sensors. Preliminary results also indicate that accuracies of 0.1 K and 1 K can be obtained below respectively 5 K and 77 K.CERN-ATS-2013-003oai:cds.cern.ch:15076102013-01-22 |
spellingShingle | Accelerators and Storage Rings Balle, C Casas, J Fortescue-Beck, E Vauthier, N Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor |
title | Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor |
title_full | Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor |
title_fullStr | Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor |
title_full_unstemmed | Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor |
title_short | Accuracy of the Approximation Function Deduced from the Fixed 3-Points Calibration Delivered with the Cernox™ Sensor |
title_sort | accuracy of the approximation function deduced from the fixed 3-points calibration delivered with the cernox™ sensor |
topic | Accelerators and Storage Rings |
url | http://cds.cern.ch/record/1507610 |
work_keys_str_mv | AT ballec accuracyoftheapproximationfunctiondeducedfromthefixed3pointscalibrationdeliveredwiththecernoxsensor AT casasj accuracyoftheapproximationfunctiondeducedfromthefixed3pointscalibrationdeliveredwiththecernoxsensor AT fortescuebecke accuracyoftheapproximationfunctiondeducedfromthefixed3pointscalibrationdeliveredwiththecernoxsensor AT vauthiern accuracyoftheapproximationfunctiondeducedfromthefixed3pointscalibrationdeliveredwiththecernoxsensor |