Magnetic shield of PMT used in DAMPE electromagnetic calorimeter
The magnetic characteristics of photomultiplier tube R5610A-01 are studied in this paper. The experimental data shows that the gain of R5610A-01 loses about 53% when the magnetic field is 3G along its +X axis. A cylinder of one-layer permalloy strip is able to reduce the effect of 3G magnetic field...