Cargando…
Thermo-dynamical measurements for ATLAS Inner Detector (evaporative cooling system)
During the construction, installation and initial operation of the Evaporative Cooling System for the ATLAS Inner Detector SCT Barrel Sub-detector, some performance characteristics were observed to be inconsistent with the original design specifications, therefore the assumptions made in the ATLAS I...
Autor principal: | Bitadze, Alexander |
---|---|
Lenguaje: | eng |
Publicado: |
2014
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/1955536 |
Ejemplares similares
-
The Evaporative Cooling System for the ATLAS Inner Detector
por: Aitree, D, et al.
Publicado: (2008) -
Development of fluorocarbon evaporative cooling recirculators and controls for the ATLAS inner silicon tracker
por: Bayer, C, et al.
Publicado: (2000) -
ATLAS Evaporative Cooling Performance
por: Wynne, B M
Publicado: (2010) -
Fluorocarbon evaporative cooling developments for the ATLAS pixel and semiconductor tracking detectors
por: Anderssen, E, et al.
Publicado: (1999) -
Evaporative cooling in ATLAS – present and future
por: Viehhauser, G
Publicado: (2010)