First on-line applications of multi-reflection time-of-flight mass separator at ISOLTRAP and the mass measurement of $^{82}$Zn
This thesis describes the implementation and first on-line application of a multi-reflection time-of-flight (MR-ToF) mass analyzer for high-resolution mass separation at the ISOLTRAP mass spectrometer at ISOLDE/CERN. On the one hand, the major objective was to improve ISOLTRAPs mass-measurement capa...
Autor principal: | |
---|---|
Lenguaje: | eng |
Publicado: |
2015
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/2063674 |
Sumario: | This thesis describes the implementation and first on-line application of a multi-reflection time-of-flight (MR-ToF) mass analyzer for high-resolution mass separation at the ISOLTRAP mass spectrometer at ISOLDE/CERN. On the one hand, the major objective was to improve ISOLTRAPs mass-measurement capabilities with respect to the ratio of delivered contaminating ions to ions of interest. On the other hand, the time necessary to purify wanted from unwanted species should be reduced as much as possible to enable access to even more exotic nuclei. The device has been set up, optimized and tested at the University of Greifswald before its move to ISOLTRAP. The achieved performance comprises mass resolving powers of up to 200000 reached at observation times of 30ms and a contamination suppression of about four orders of magnitude by use of a Bradbury-Nielsen gate. With the characteristics, it outperforms clearly the so far state-of-the-art purification method of a gas-filled Penning trap. To improve the utilization of the MR-ToF mass analyzer, the in-trap lift method has been developed. It simplifies the application and optimization of the device, which is a crucial time factor in an on-line experiment. The device was the first of its kind successfully applied to radioactive ion beams for a mass analysis, for a mass separation (in combination with the Bradbury-Nielsen gate) as a preparatory step for a subsequent Penning-trap mass measurement and as a high-precision mass spectrometer of its own. The latter was recently used for the first mass measurement of the neutron-rich calcium isotopes $^{53}$Ca and $^{54}$Ca. The so-far achieved mass-resolving power of 200000 belongs to the highest reported for time-of-flight mass analyzers at all. The first successful application of the MR-ToF system as the only mass separator at ISOLTRAP resulted in the mass measurement of $^{82}$Zn. The new mass value has been compared to mass extrapolations of the most recent Hartree-Fock-Bogoliubov (HFB) mass models, HFB-19 to HFB-21, of the BRUSLIB collaboration. The mass of the nuclide is of high interest for the compositions and depth profile of the outer crust of neutron stars. In the classical model of the outer crust of a cold, non-accreting and non-rotating neutron star, the sequence of nuclides found within this parts is determined mainly by the binding energy of exotic nuclides. The crustal compositions determined with the three HFB mass models differed with respect to the appearance of a layer of $^{82}$Zn, originating from different mass extrapolations of this mass. With the new experimental data, the extrapolations could be evaluated. It was found that the HFB-21 mass value differs less from the experimental data than the ones from HFB-19 and 20. Therefore, in the classical model, $^{82}$Zn does not appear anymore in the outer crust. Due to its high resolution and very fast measurement time, the MR-ToF mass analyzer will be an important instruments for future activities at ISOLTRAP, at the ISOLDE facility in general, and at other radioactive ion-beam facilities. |
---|