Rotating-coil calibration in a reference quadrupole, considering roll-angle misalignment and higher-order harmonics
A method is proposed for calibrating the radius of a rotating coil sensor by relaxing the metrological constraints on alignment and field errors of the reference quadrupole. A coil radius calibration considering a roll-angle misalignment of the measurement bench, the magnet, and the motor-drive unit...
Autores principales: | , , , , |
---|---|
Formato: | info:eu-repo/semantics/article |
Lenguaje: | eng |
Publicado: |
Measurement
2016
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1016/j.measurement.2016.02.061 http://cds.cern.ch/record/2148402 |
Sumario: | A method is proposed for calibrating the radius of a rotating coil sensor by relaxing the metrological constraints on alignment and field errors of the reference quadrupole. A coil radius calibration considering a roll-angle misalignment of the measurement bench, the magnet, and the motor-drive unit is analyzed. Then, the error arising from higher-order harmonic field imperfections in the reference quadrupole is assessed. The method is validated by numerical field computation for both the higher-order harmonic errors and the roll-angle misalignment. Finally, an experimental proof-of-principle demonstration is car-ried out in a calibration magnet with sextupole harmonic. |
---|