First performance evaluation of a Multi-layer Thick Gaseous Electron Multiplier with in-built electrode meshes - MM-THGEM
We describe a new micro-pattern gas detector structure comprising a multi-layer hole-type multiplier (M-THGEM) combined with two in-built electrode meshes: the Multi-Mesh THGEM-type multiplier (MM-THGEM). Suitable potential differences applied between the various electrodes provide an efficient coll...
Autores principales: | , |
---|---|
Lenguaje: | eng |
Publicado: |
2018
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1088/1748-0221/13/06/P06019 http://cds.cern.ch/record/2317339 |
Sumario: | We describe a new micro-pattern gas detector structure comprising a multi-layer hole-type multiplier (M-THGEM) combined with two in-built electrode meshes: the Multi-Mesh THGEM-type multiplier (MM-THGEM). Suitable potential differences applied between the various electrodes provide an efficient collection of ionization electrons within the MM-THGEM holes and a large charge avalanche multiplication between the meshes. Different from conventional hole-type multipliers (e.g. Gas Electron Multipliers - GEMs, Thick Gas Electron Multipliers—THGEMs, etc.), which are characterized by a variable (dipole-like) field strength inside the avalanche gap, electrons in MM-THGEMs are largely multiplied by a strong uniform field established between the two meshes, like in the parallel-plate avalanche geometry. The presence of the two meshes within the holes allows for the trapping of a large fraction of the positive ions that stream back to the drift region. A gas gain above 105 has been achieved for single photo-electron detection with a single MM-THGEM in Ar/(10%)CH4 and He/(10%)CO2, at standard conditions for temperature and pressure. When the MM-THGEM is coupled to a conventional THGEM and used as first cascade element, the maximum achievable gains reach values above 106 in He/(10%)CO2, while the IBF approaches of 1.5% in the case of optimum detector-bias configuration. This IBF value is several times lower compared to the one obtained by a double GEM/THGEM detector (5–10%), and equivalent to the performance attained by a Micromegas detector. |
---|