Personal Dosimetry in Direct Pulsed Photon Fields with the Dosepix Detector
First investigations regarding dosimetric properties of the hybrid, pixelated, photon-counting Dosepix detector in the direct beam of a pulsed photon field (RQR8) for the personal dose equivalent H p (10) are presented. The influence quantities such as pulse duration and dose rate were varied, and...
Autores principales: | , , , , , , , , , , , , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2021
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1109/TNS.2022.3222544 http://cds.cern.ch/record/2773259 |
Sumario: | First investigations regarding dosimetric properties of the hybrid, pixelated, photon-counting Dosepix detector in the direct beam of a pulsed photon field (RQR8) for the personal dose equivalent
H
p
(10)
are presented. The influence quantities such as pulse duration and dose rate were varied, and their responses were compared to the legal limits provided in PTB-A 23.2. The variation of pulse duration at a nearly constant dose rate of about 3.7 Sv/h shows a flat response around 1.0 from 3.6 s down to 2 ms. A response close to 1.0 is achieved for dose rates from 0.07 to 35 Sv/h for both pixel sizes. Above this dose rate, the large pixels (
220 μm
edge length) are below the lower limit. The small pixels (
55 μm
edge length) stay within limits up to 704 Sv/h. The count rate linearity is compared to previous results, confirming the saturating count rate for high dose rates. |
---|