Cargando…
Impact of the newly proposed ICRU/ICRP quantities on neutron calibration fields and extended range neutron rem-counters
In July 2017, the International Commission on Radiation Units and Measurements (ICRU) and the International Commission on Radiological Protection (ICRP) proposed the introduction of new operational quantities for external radiation exposure, with the aim of improving coherence between protection qua...
Autores principales: | , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2019
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1088/1361-6498/ab18ca http://cds.cern.ch/record/2801577 |
_version_ | 1780972706337914880 |
---|---|
author | Pozzi, Fabio Ferrarini, Michele Ferrulli, Francesca Silari, Marco |
author_facet | Pozzi, Fabio Ferrarini, Michele Ferrulli, Francesca Silari, Marco |
author_sort | Pozzi, Fabio |
collection | CERN |
description | In July 2017, the International Commission on Radiation Units and Measurements (ICRU) and the International Commission on Radiological Protection (ICRP) proposed the introduction of new operational quantities for
external radiation exposure, with the aim of improving coherence between
protection quantities and operational quantities within the system of radiological protection. A change in operational quantities will impact both
instrumentation and reference radiation fields used for their calibration. This
paper evaluates the potential impact of the new quantity ambient dose, $H$*
,
meant to replace ambient dose equivalent, $H$*
(10), on two neutron reference
fields, the Am-Be source and the CERF high-energy workplace field, and on
the response of two models of extended-range neutron rem counters (LINUS
and LUPIN). The conclusions are that calibration procedures should in general
not be affected and that changes should only be expected in calibration
coefficients. Considering the acceptable measurement uncertainties for
operational radiation protection, for the extended-range rem counters changes
in their design would not be required for measurements outside particle
accelerators shielding and for aircrew dosimetry. One can expect that this type
of instrument can still be calibrated with Am-Be source neutrons and
employed in neutron fields with energy distributions spanning several decades.
For uses in radiation fields with very peculiar neutron energy distribution, a
specific workplace field calibration may instead be required. |
id | cern-2801577 |
institution | Organización Europea para la Investigación Nuclear |
language | eng |
publishDate | 2019 |
record_format | invenio |
spelling | cern-28015772022-10-28T12:31:03Zdoi:10.1088/1361-6498/ab18cahttp://cds.cern.ch/record/2801577engPozzi, FabioFerrarini, MicheleFerrulli, FrancescaSilari, MarcoImpact of the newly proposed ICRU/ICRP quantities on neutron calibration fields and extended range neutron rem-countersNuclear Physics - ExperimentIn July 2017, the International Commission on Radiation Units and Measurements (ICRU) and the International Commission on Radiological Protection (ICRP) proposed the introduction of new operational quantities for external radiation exposure, with the aim of improving coherence between protection quantities and operational quantities within the system of radiological protection. A change in operational quantities will impact both instrumentation and reference radiation fields used for their calibration. This paper evaluates the potential impact of the new quantity ambient dose, $H$* , meant to replace ambient dose equivalent, $H$* (10), on two neutron reference fields, the Am-Be source and the CERF high-energy workplace field, and on the response of two models of extended-range neutron rem counters (LINUS and LUPIN). The conclusions are that calibration procedures should in general not be affected and that changes should only be expected in calibration coefficients. Considering the acceptable measurement uncertainties for operational radiation protection, for the extended-range rem counters changes in their design would not be required for measurements outside particle accelerators shielding and for aircrew dosimetry. One can expect that this type of instrument can still be calibrated with Am-Be source neutrons and employed in neutron fields with energy distributions spanning several decades. For uses in radiation fields with very peculiar neutron energy distribution, a specific workplace field calibration may instead be required.oai:cds.cern.ch:28015772019 |
spellingShingle | Nuclear Physics - Experiment Pozzi, Fabio Ferrarini, Michele Ferrulli, Francesca Silari, Marco Impact of the newly proposed ICRU/ICRP quantities on neutron calibration fields and extended range neutron rem-counters |
title | Impact of the newly proposed ICRU/ICRP quantities on neutron calibration fields and extended range neutron rem-counters |
title_full | Impact of the newly proposed ICRU/ICRP quantities on neutron calibration fields and extended range neutron rem-counters |
title_fullStr | Impact of the newly proposed ICRU/ICRP quantities on neutron calibration fields and extended range neutron rem-counters |
title_full_unstemmed | Impact of the newly proposed ICRU/ICRP quantities on neutron calibration fields and extended range neutron rem-counters |
title_short | Impact of the newly proposed ICRU/ICRP quantities on neutron calibration fields and extended range neutron rem-counters |
title_sort | impact of the newly proposed icru/icrp quantities on neutron calibration fields and extended range neutron rem-counters |
topic | Nuclear Physics - Experiment |
url | https://dx.doi.org/10.1088/1361-6498/ab18ca http://cds.cern.ch/record/2801577 |
work_keys_str_mv | AT pozzifabio impactofthenewlyproposedicruicrpquantitiesonneutroncalibrationfieldsandextendedrangeneutronremcounters AT ferrarinimichele impactofthenewlyproposedicruicrpquantitiesonneutroncalibrationfieldsandextendedrangeneutronremcounters AT ferrullifrancesca impactofthenewlyproposedicruicrpquantitiesonneutroncalibrationfieldsandextendedrangeneutronremcounters AT silarimarco impactofthenewlyproposedicruicrpquantitiesonneutroncalibrationfieldsandextendedrangeneutronremcounters |