Testing the Properties of Beam-Dose Monitors for VHEE-FLASH Radiation Therapy

Very High Energy Electrons (VHEE) of 50 - 250 MeV are an attractive choice for FLASH radiation therapy (RT). Before VHEE-FLASH RT can be considered for clinical use, a reliable dosimetric and beam monitoring system needs to be developed, able to measure the dose delivered to the patient in real-time...

Descripción completa

Detalles Bibliográficos
Autores principales: Bateman, Joseph, Burrows, Philip, Corsini, Roberto, Dosanjh, Manjit, Dyks, Luke, Farabolini, Wilfrid, Gerbershagen, Alexander, Heracleous, Natalie, Korysko, Pierre, Morales Vigo, Sara, Murtas, Fabrizio, Rieker, Vilde, Salvachúa, Belen, Silari, Marco, Zorloni, Gabriele
Lenguaje:eng
Publicado: 2022
Materias:
Acceso en línea:https://dx.doi.org/10.18429/JACoW-IPAC2022-THPOMS029
http://cds.cern.ch/record/2845863
Descripción
Sumario:Very High Energy Electrons (VHEE) of 50 - 250 MeV are an attractive choice for FLASH radiation therapy (RT). Before VHEE-FLASH RT can be considered for clinical use, a reliable dosimetric and beam monitoring system needs to be developed, able to measure the dose delivered to the patient in real-time and cut off the beam in the event of a machine fault to prevent overdosing the patient. Ionisation chambers are the standard monitors in conventional RT; however, their response saturates at the high dose rates required for FLASH. Therefore, a new dosimetry method is needed that can provide reliable measurements of the delivered dose in these conditions. Experiments using 200 MeV electrons were done at the CLEAR facility at CERN to investigate the properties of detectors such as diamond beam loss detectors, GEM foil detectors, and Timepix3 ASIC chips. From the tests, the GEM foil proved to be the most promising.