Beam tests of lead tungstate crystal matrices and a silicon strip preshower detector for the CMS electromagnetic calorimeter

Tests of lead tungstate crystal matrices carried out in high-energy electron beams in 1996, using new crystals, new APDs and an improved test set-up, confirm that an energy resolution of better than 0 .6% at 100 GeV can be obtained when the longitudinal uniformity of the struck crystal is adequate....

Descripción completa

Detalles Bibliográficos
Autores principales: Auffray, Etiennette, Baillon, Paul, Barney, D, Bassompierre, Gabriel, Benhammou, Ya, Blick, A M, Bloch, P, Bonamy, P, Bourotte, J, Buiron, L, Cavallari, F, Chipaux, Rémi, Cockerill, D J A, Dafinei, I, Davies, G, Depasse, P, Deiters, K, Diemoz, M, Dobrzynski, Ludwik, Donskov, S V, Mamouni, H E, Ercoli, C, Faure, J L, Felcini, Marta, Gautheron, F, Géléoc, M, Givernaud, Alain, Gninenko, S N, Godinovic, N, Graham, D J, Guillaud, J P, Guschin, E, Haguenauer, Maurice, Hillemanns, H, Hofer, H, Ille, B, Inyakin, A V, Jääskeläinen, S, Katchanov, V A, Kirn, T, Kloukinas, Kostas C, Korzhik, M V, Lassila-Perini, K M, Lebrun, P, Lecoq, P, Lecoeur, Gérard, Lecomte, P, Leonardi, E, Locci, E, Loos, R, Longo, E, MacKay, C K, Martin, E, Mendiburu, J P, Musienko, Yu V, Nédélec, P, Nessi-Tedaldi, F, Organtini, G, Paoletti, S, Pansart, J P, Peigneux, J P, Puljak, I, Qian, S, Reid, E, Renker, D, Rosowsky, A, Rosso, E, Rusack, R W, Rykaczewski, H, Schneegans, M, Seez, Christopher J, Semeniouk, I N, Shagin, P M, Sillou, D, Singovsky, A V, Sougonyaev, V, Soric, I, Verrecchia, P, Vialle, J P, Virdee, Tejinder S, Zhu, R Y
Lenguaje:eng
Publicado: 1998
Materias:
Acceso en línea:https://dx.doi.org/10.1016/S0168-9002(98)00464-1
http://cds.cern.ch/record/351240
Descripción
Sumario:Tests of lead tungstate crystal matrices carried out in high-energy electron beams in 1996, using new crystals, new APDs and an improved test set-up, confirm that an energy resolution of better than 0 .6% at 100 GeV can be obtained when the longitudinal uniformity of the struck crystal is adequate. Light loss measurements under low dose irradiation are reported. It is shown that there is no loss of energy resolution after irradiation and that the calibration change due to light loss can be tracked with a precision monitoring system. Finally, successuful tests with a preshower device, equipped wi th silicon strip detector readout, are described.