Breakdown Limit Studies in High Rate Gaseous Detectors
We report results from a systematic study of breakdown limits for novel high rate gaseous detectors: MICROMEGAS, CAT and GEM, together with more conventional devices such as thin-gap parallel-mesh chambers and high-rate wire chambers. It was found that for all these detectors, the maximum achievable...
Autores principales: | Ivaniouchenkov, Yu, Fonte, Paulo J R, Peskov, Vladimir, Ramsey, B D |
---|---|
Lenguaje: | eng |
Publicado: |
1998
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/367263 |