Fourth intercomparison of personal dosemeters used in US Department of Energy accelerator facilities

Personal neutron dosemeters from seven US Department of Energy (DOE) laboratories were mailed to the European Laboratory for Particle Physics (CERN) and irradiated using the well-characterised CERN reference radiation facility (CERF). Neutron dose equivalents determined using the DOE personal doseme...

Descripción completa

Detalles Bibliográficos
Autores principales: Stewart, R D, McDonald, J C, Otto, T, Loesch, R M
Lenguaje:eng
Publicado: 2000
Materias:
Acceso en línea:http://cds.cern.ch/record/430807
Descripción
Sumario:Personal neutron dosemeters from seven US Department of Energy (DOE) laboratories were mailed to the European Laboratory for Particle Physics (CERN) and irradiated using the well-characterised CERN reference radiation facility (CERF). Neutron dose equivalents determined using the DOE personal dosemeters have been compared to the reference dose equivalent as determined using a tissue-equivalent proportional counter (TEPC). In the 0.5 to 5 mSv dose equivalent range, the comparison of results suggests that the neutron personal dosemeters in use at DOE facilities are capable of estimating dose equivalents for high energy neutrons to within a factor of at least 2 or 3. If a field-specific calibration factor is used to correct the dose equivalent responses, the agreement with the reference dose equivalent for these dosemeters can be improved to better than about 25 to 65at is decoupled from the core in /sup 183,185/Au, becomes the 3/2[532] state (h9/2 parentage) strongly coupled in the doubly-odd /sup 184/Au nucleus. (10 refs).