In this paper Monte Carlo simulations (FLUKA) and measurements of the response of a BGO detector are reported. %For the measurements different radioactive sources were used to irradiate the BGO crystal. For the measurements three low-energy photon emitters $\left({}^{60}\rm{Co},\right.$ ${}^{54}\rm{...

Descripción completa

Detalles Bibliográficos
Autores principales: Vincke, H H, Gschwendtner, E, Fabjan, Christian Wolfgang, Otto, T
Lenguaje:eng
Publicado: 2001
Materias:
Acceso en línea:https://dx.doi.org/10.1016/S0168-9002(01)01966-0
http://cds.cern.ch/record/507785
Descripción
Sumario:In this paper Monte Carlo simulations (FLUKA) and measurements of the response of a BGO detector are reported. %For the measurements different radioactive sources were used to irradiate the BGO crystal. For the measurements three low-energy photon emitters $\left({}^{60}\rm{Co},\right.$ ${}^{54}\rm{Mn},$ $\left. {}^{137}\rm{Cs}\right)$ were used to irradiate the BGO from various distances and angles. The neutron response was measured with an Am--Be neutron source. Simulations of the experimental irradiations were carried out. Our study can also be considered as a benchmark for FLUKA in terms of its reliability to predict the detector response of a BGO scintillator.