Response of a BGO detector to photon and neutron sources: simulations and measurements
In this paper Monte Carlo simulations (FLUKA) and measurements of the response of a BGO detector are reported. %For the measurements different radioactive sources were used to irradiate the BGO crystal. For the measurements three low-energy photon emitters $\left({}^{60}\rm{Co},\right.$ ${}^{54}\rm{...
Autores principales: | , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2001
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1016/S0168-9002(01)01966-0 http://cds.cern.ch/record/507785 |
Sumario: | In this paper Monte Carlo simulations (FLUKA) and measurements of the response of a BGO detector are reported. %For the measurements different radioactive sources were used to irradiate the BGO crystal. For the measurements three low-energy photon emitters $\left({}^{60}\rm{Co},\right.$ ${}^{54}\rm{Mn},$ $\left. {}^{137}\rm{Cs}\right)$ were used to irradiate the BGO from various distances and angles. The neutron response was measured with an Am--Be neutron source. Simulations of the experimental irradiations were carried out. Our study can also be considered as a benchmark for FLUKA in terms of its reliability to predict the detector response of a BGO scintillator. |
---|