The response of the Sievert instrument in neutron beams up to 180 MeV

Measurements with a tissue-equivalent proportional counter (TEPC) using the variance-covariance method have been performed in neutron beams between 71 keV and 180 MeV and in the cosmic radiation reference field (CERF) at CERN. The results show that with appropriate linear Q/sub D/(y/sub D/) relation...

Descripción completa

Detalles Bibliográficos
Autores principales: Kylionen, J E, Lindborg, L, Samuelson, G
Lenguaje:eng
Publicado: 2001
Materias:
Acceso en línea:http://cds.cern.ch/record/516160
_version_ 1780897623805263872
author Kylionen, J E
Lindborg, L
Samuelson, G
author_facet Kylionen, J E
Lindborg, L
Samuelson, G
author_sort Kylionen, J E
collection CERN
description Measurements with a tissue-equivalent proportional counter (TEPC) using the variance-covariance method have been performed in neutron beams between 71 keV and 180 MeV and in the cosmic radiation reference field (CERF) at CERN. The results show that with appropriate linear Q/sub D/(y/sub D/) relations, the ambient dose equivalent can be determined within about 55% in these beams. Build- up measurements show that wall thickness is not crucial for H* determinations at 60 and 180 MeV. (26 refs).
id cern-516160
institution Organización Europea para la Investigación Nuclear
language eng
publishDate 2001
record_format invenio
spelling cern-5161602019-09-30T06:29:59Zhttp://cds.cern.ch/record/516160engKylionen, J ELindborg, LSamuelson, GThe response of the Sievert instrument in neutron beams up to 180 MeVHealth Physics and Radiation EffectsMeasurements with a tissue-equivalent proportional counter (TEPC) using the variance-covariance method have been performed in neutron beams between 71 keV and 180 MeV and in the cosmic radiation reference field (CERF) at CERN. The results show that with appropriate linear Q/sub D/(y/sub D/) relations, the ambient dose equivalent can be determined within about 55% in these beams. Build- up measurements show that wall thickness is not crucial for H* determinations at 60 and 180 MeV. (26 refs).oai:cds.cern.ch:5161602001
spellingShingle Health Physics and Radiation Effects
Kylionen, J E
Lindborg, L
Samuelson, G
The response of the Sievert instrument in neutron beams up to 180 MeV
title The response of the Sievert instrument in neutron beams up to 180 MeV
title_full The response of the Sievert instrument in neutron beams up to 180 MeV
title_fullStr The response of the Sievert instrument in neutron beams up to 180 MeV
title_full_unstemmed The response of the Sievert instrument in neutron beams up to 180 MeV
title_short The response of the Sievert instrument in neutron beams up to 180 MeV
title_sort response of the sievert instrument in neutron beams up to 180 mev
topic Health Physics and Radiation Effects
url http://cds.cern.ch/record/516160
work_keys_str_mv AT kylionenje theresponseofthesievertinstrumentinneutronbeamsupto180mev
AT lindborgl theresponseofthesievertinstrumentinneutronbeamsupto180mev
AT samuelsong theresponseofthesievertinstrumentinneutronbeamsupto180mev
AT kylionenje responseofthesievertinstrumentinneutronbeamsupto180mev
AT lindborgl responseofthesievertinstrumentinneutronbeamsupto180mev
AT samuelsong responseofthesievertinstrumentinneutronbeamsupto180mev