Development of a Mass Flowmeter based on the Coriolis Acceleration for Liquid, Supercritical and Superfluid Helium

Beginning in the 1980's, Coriolis meters have gained generalised acceptance in liquid applications with a worldwide installed base of over 300,000 units. To meet the demands of cryogenic applications below 20 K, off-the-shelf Coriolis meters have been used, with minor design modifications and o...

Descripción completa

Detalles Bibliográficos
Autores principales: De Jonge, T, Patten, T, Rivetti, A, Serio, L
Lenguaje:eng
Publicado: 2002
Materias:
Acceso en línea:http://cds.cern.ch/record/593503
Descripción
Sumario:Beginning in the 1980's, Coriolis meters have gained generalised acceptance in liquid applications with a worldwide installed base of over 300,000 units. To meet the demands of cryogenic applications below 20 K, off-the-shelf Coriolis meters have been used, with minor design modifications and operational changes. The meters were originally calibrated on water and tested on liquid helium at 4.5 K, supercritical helium around 5 K and superfluid helium below 2 K. The meters maintain their intrinsic robustness and accuracy of better than 1% of measured value; accuracy is independent of density and temperature.