Carbon clusters for absolute mass measurements at ISOLTRAP

The cyclotron frequencies of singly charged carbon clusters C/sub n //sup +/ (n >or= 2) were measured with the Penning-trap mass spectrometer ISOLTRAP at ISOLDE/CERN. The present limit of mass accuracy delta m/m = 1.2.10/sup -8/ and the extent of the mass- dependent systematic shift ( delta m/m)/...

Descripción completa

Detalles Bibliográficos
Autores principales: Blaum, K, Bollen, G, Herfurth, F, Kellerbauer, A G, Kluge, H J, Kuckein, M, Sauvan, E, Scheidenberger, C, Schweikhard, L
Lenguaje:eng
Publicado: 2002
Materias:
Acceso en línea:https://dx.doi.org/10.1140/epja/i2001-10262-4
http://cds.cern.ch/record/623217
Descripción
Sumario:The cyclotron frequencies of singly charged carbon clusters C/sub n //sup +/ (n >or= 2) were measured with the Penning-trap mass spectrometer ISOLTRAP at ISOLDE/CERN. The present limit of mass accuracy delta m/m = 1.2.10/sup -8/ and the extent of the mass- dependent systematic shift ( delta m/m)/sub sys/ = 1.7(0.6).10/sup -10//u.(m - m/sub ref/) of the setup were investigated for the first time. In addition, absolute mass measurements by use of pure clusters of the most abundant carbon isotope /sup 12/C are now possible at ISOLTRAP. (15 refs).