Test of the rem-counter WENDI-II from Eberline in different energy-dispersed neutron fields
The neutron rem-counter WENDI-II from Eberline was tested in high-energy particle accelerator produced neutron fields. A radioactive 241Am-Be(α,n) source was used as a reference. The experimentally determined responses are compared to Monte-Carlo simulations of the response function done by Olsher e...