Type testing of the SDM2000 personal dosimeter

The results of the performance of a new type of personal dosimeter, the SDM2000 Personal Dosimeter made by China Institute for Radiation Protection is presented. Tests were performed on radiological performance including radiation energy, incidence angular, beta radiation, neutron radiation, accurac...

Descripción completa

Detalles Bibliográficos
Autores principales: LiuZhengShan, Cheng Chang, Deng Chang Ming, Guo Zhan Jie, Zhang Zhi Yong
Lenguaje:chi
Publicado: 2001
Materias:
Acceso en línea:http://cds.cern.ch/record/748225
_version_ 1780904232948334592
author LiuZhengShan
Cheng Chang
Deng Chang Ming
Guo Zhan Jie
Zhang Zhi Yong
author_facet LiuZhengShan
Cheng Chang
Deng Chang Ming
Guo Zhan Jie
Zhang Zhi Yong
author_sort LiuZhengShan
collection CERN
description The results of the performance of a new type of personal dosimeter, the SDM2000 Personal Dosimeter made by China Institute for Radiation Protection is presented. Tests were performed on radiological performance including radiation energy, incidence angular, beta radiation, neutron radiation, accuracy of alarm levels, relative intrinsic error, overload, response time, linearity, retention of reading. There are also tests on the effects of a variety of environmental factors, such as temperature, humidity, external magnetic fields, electrostatic discharge, external electromagnetic fields. Other characteristics tested were dropping test, vibration test and battery life. The test results were compared with the relevant requirements of three standards: CIRP standard: Q/DSC1-92, National standard: GB/T 13161-91, IEC standard: IEC1283. In general, the performance of the SDM2000 personal dosimeter was found to be quite acceptable; it meets most of the relevant requirements of the three standards
id cern-748225
institution Organización Europea para la Investigación Nuclear
language chi
publishDate 2001
record_format invenio
spelling cern-7482252019-09-30T06:29:59Zhttp://cds.cern.ch/record/748225chiLiuZhengShanCheng ChangDeng Chang MingGuo Zhan JieZhang Zhi YongType testing of the SDM2000 personal dosimeterDetectors and Experimental TechniquesThe results of the performance of a new type of personal dosimeter, the SDM2000 Personal Dosimeter made by China Institute for Radiation Protection is presented. Tests were performed on radiological performance including radiation energy, incidence angular, beta radiation, neutron radiation, accuracy of alarm levels, relative intrinsic error, overload, response time, linearity, retention of reading. There are also tests on the effects of a variety of environmental factors, such as temperature, humidity, external magnetic fields, electrostatic discharge, external electromagnetic fields. Other characteristics tested were dropping test, vibration test and battery life. The test results were compared with the relevant requirements of three standards: CIRP standard: Q/DSC1-92, National standard: GB/T 13161-91, IEC standard: IEC1283. In general, the performance of the SDM2000 personal dosimeter was found to be quite acceptable; it meets most of the relevant requirements of the three standardsCIRP-0034CNIC-01581oai:cds.cern.ch:7482252001
spellingShingle Detectors and Experimental Techniques
LiuZhengShan
Cheng Chang
Deng Chang Ming
Guo Zhan Jie
Zhang Zhi Yong
Type testing of the SDM2000 personal dosimeter
title Type testing of the SDM2000 personal dosimeter
title_full Type testing of the SDM2000 personal dosimeter
title_fullStr Type testing of the SDM2000 personal dosimeter
title_full_unstemmed Type testing of the SDM2000 personal dosimeter
title_short Type testing of the SDM2000 personal dosimeter
title_sort type testing of the sdm2000 personal dosimeter
topic Detectors and Experimental Techniques
url http://cds.cern.ch/record/748225
work_keys_str_mv AT liuzhengshan typetestingofthesdm2000personaldosimeter
AT chengchang typetestingofthesdm2000personaldosimeter
AT dengchangming typetestingofthesdm2000personaldosimeter
AT guozhanjie typetestingofthesdm2000personaldosimeter
AT zhangzhiyong typetestingofthesdm2000personaldosimeter