Dosimetric response of united, commercially available CTA foils for sup 6 sup 0 Co gamma rays

The usefulness of two kinds of untinted CTA foils: Fuji CTR-125 dosimetric foil and technical CTA-T foil, produced by 'Zaklady Chemiczne, 'Gorzow Wielkopolski' as support for light-sensitive layers of amateur photo-films, for sup 6 sup 0 Co gamma ray dosimetry was investigated. In spi...

Descripción completa

Detalles Bibliográficos
Autor principal: Peimel-Stuglik, Z
Lenguaje:pol
Publicado: 2001
Materias:
Acceso en línea:http://cds.cern.ch/record/748320
Descripción
Sumario:The usefulness of two kinds of untinted CTA foils: Fuji CTR-125 dosimetric foil and technical CTA-T foil, produced by 'Zaklady Chemiczne, 'Gorzow Wielkopolski' as support for light-sensitive layers of amateur photo-films, for sup 6 sup 0 Co gamma ray dosimetry was investigated. In spite of rather bad physical parameters of the technical foil (spread of foil thickness, high and different initial absorbance) the dosimetric response of both foils for sup 6 sup 0 Co gamma rays was similar. The CTA-T foil can be used for routine dosimetry providing that dosimetric signals have to be calculated exactly as recommended by the ASTM (American Society for Testing and Materials) standard, i.e. as the difference of absorbance of irradiated and (the same) non-irradiated foil. Any other approach may lead to high errors of dose evaluation. The last is true also for other CTA foils, especially after long self-life.