Cargando…
Response of RadFET Dosimeters to High Fluences of Fast Neutrons
We present irradiation experiments carried out on RadFETs in the 20.4 MeV high intensity T2 neutron beam at the CRC-UCL in Belgium. The aim of the test was to characterize the neutron response of RadFETs in view of their use as an integrated part of a radiationmonitoring sensor for the CERN Large Ha...
Autores principales: | , , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2004
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1109/TNS.2005.852709 http://cds.cern.ch/record/798311 |
_version_ | 1780904706237792256 |
---|---|
author | Ravotti, F Glaser, M Moll, Michael Ilgner, C Camanzi, B Holmes-Siedle, A G |
author_facet | Ravotti, F Glaser, M Moll, Michael Ilgner, C Camanzi, B Holmes-Siedle, A G |
author_sort | Ravotti, F |
collection | CERN |
description | We present irradiation experiments carried out on RadFETs in the 20.4 MeV high intensity T2 neutron beam at the CRC-UCL in Belgium. The aim of the test was to characterize the neutron response of RadFETs in view of their use as an integrated part of a radiationmonitoring sensor for the CERN Large Hadron Collider (LHC) experiments. Two types of RadFETs were investigated up to a total neutron fluence of 3x10^14 cm-2 corresponding to a deposited dose of 744 Gy in Silicon. The responses of bare devices to neutrons are compared to the commonly used reference measurements with gamma rays. It is found that the gamma ray calibration cannot directly be adopted to convert the RadFET signals into neutron dose. In a second experiment, the influence of a plastic packaging, simulated by Polyethylene slabs of different thicknesses, was tested in the neutron beam and compared to GEANT4 Monte Carlo simulations. An increase of the RadFET neutron sensitivity by a factor of up to 7 due to the packaging is found. The influence of these findings on the conception of the radiationmonitoring sensor is discussed. |
id | cern-798311 |
institution | Organización Europea para la Investigación Nuclear |
language | eng |
publishDate | 2004 |
record_format | invenio |
spelling | cern-7983112019-09-30T06:29:59Zdoi:10.1109/TNS.2005.852709http://cds.cern.ch/record/798311engRavotti, FGlaser, MMoll, MichaelIlgner, CCamanzi, BHolmes-Siedle, A GResponse of RadFET Dosimeters to High Fluences of Fast NeutronsHealth Physics and Radiation EffectsDetectors and Experimental TechniquesWe present irradiation experiments carried out on RadFETs in the 20.4 MeV high intensity T2 neutron beam at the CRC-UCL in Belgium. The aim of the test was to characterize the neutron response of RadFETs in view of their use as an integrated part of a radiationmonitoring sensor for the CERN Large Hadron Collider (LHC) experiments. Two types of RadFETs were investigated up to a total neutron fluence of 3x10^14 cm-2 corresponding to a deposited dose of 744 Gy in Silicon. The responses of bare devices to neutrons are compared to the commonly used reference measurements with gamma rays. It is found that the gamma ray calibration cannot directly be adopted to convert the RadFET signals into neutron dose. In a second experiment, the influence of a plastic packaging, simulated by Polyethylene slabs of different thicknesses, was tested in the neutron beam and compared to GEANT4 Monte Carlo simulations. An increase of the RadFET neutron sensitivity by a factor of up to 7 due to the packaging is found. The influence of these findings on the conception of the radiationmonitoring sensor is discussed.CERN-PH-EP-2004-045oai:cds.cern.ch:7983112004-08-31 |
spellingShingle | Health Physics and Radiation Effects Detectors and Experimental Techniques Ravotti, F Glaser, M Moll, Michael Ilgner, C Camanzi, B Holmes-Siedle, A G Response of RadFET Dosimeters to High Fluences of Fast Neutrons |
title | Response of RadFET Dosimeters to High Fluences of Fast Neutrons |
title_full | Response of RadFET Dosimeters to High Fluences of Fast Neutrons |
title_fullStr | Response of RadFET Dosimeters to High Fluences of Fast Neutrons |
title_full_unstemmed | Response of RadFET Dosimeters to High Fluences of Fast Neutrons |
title_short | Response of RadFET Dosimeters to High Fluences of Fast Neutrons |
title_sort | response of radfet dosimeters to high fluences of fast neutrons |
topic | Health Physics and Radiation Effects Detectors and Experimental Techniques |
url | https://dx.doi.org/10.1109/TNS.2005.852709 http://cds.cern.ch/record/798311 |
work_keys_str_mv | AT ravottif responseofradfetdosimeterstohighfluencesoffastneutrons AT glaserm responseofradfetdosimeterstohighfluencesoffastneutrons AT mollmichael responseofradfetdosimeterstohighfluencesoffastneutrons AT ilgnerc responseofradfetdosimeterstohighfluencesoffastneutrons AT camanzib responseofradfetdosimeterstohighfluencesoffastneutrons AT holmessiedleag responseofradfetdosimeterstohighfluencesoffastneutrons |