Cargando…
The Calibration Stategy of CMS Electromagnetic Calorimeter
Calibration is one of the main factors that set limits on the ultimate performance of the CMS electromagnetic calorimeter (ECAL) at LHC. Crystals raw intercalibration from laboratory measurements during assembly will be available for all the crystals and has been demonstrated to be a good precalibra...
Autor principal: | |
---|---|
Lenguaje: | eng |
Publicado: |
2004
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/810667 |
_version_ | 1780905294547648512 |
---|---|
author | Meridiani, P |
author_facet | Meridiani, P |
author_sort | Meridiani, P |
collection | CERN |
description | Calibration is one of the main factors that set limits on the ultimate performance of the CMS electromagnetic calorimeter (ECAL) at LHC. Crystals raw intercalibration from laboratory measurements during assembly will be available for all the crystals and has been demonstrated to be a good precalibration value at the start-up; some crystals will be also intercalibrated using an electron beam. In situ calibration with physics events will be the main tool to reduce the constant term of the emergy resolution to the design goal of 0.5%. In the following the calibration strategy will be described in detail. |
id | cern-810667 |
institution | Organización Europea para la Investigación Nuclear |
language | eng |
publishDate | 2004 |
record_format | invenio |
spelling | cern-8106672019-09-30T06:29:59Zhttp://cds.cern.ch/record/810667engMeridiani, PThe Calibration Stategy of CMS Electromagnetic CalorimeterDetectors and Experimental TechniquesCalibration is one of the main factors that set limits on the ultimate performance of the CMS electromagnetic calorimeter (ECAL) at LHC. Crystals raw intercalibration from laboratory measurements during assembly will be available for all the crystals and has been demonstrated to be a good precalibration value at the start-up; some crystals will be also intercalibrated using an electron beam. In situ calibration with physics events will be the main tool to reduce the constant term of the emergy resolution to the design goal of 0.5%. In the following the calibration strategy will be described in detail.CMS-CR-2004-045oai:cds.cern.ch:8106672004-10-01 |
spellingShingle | Detectors and Experimental Techniques Meridiani, P The Calibration Stategy of CMS Electromagnetic Calorimeter |
title | The Calibration Stategy of CMS Electromagnetic Calorimeter |
title_full | The Calibration Stategy of CMS Electromagnetic Calorimeter |
title_fullStr | The Calibration Stategy of CMS Electromagnetic Calorimeter |
title_full_unstemmed | The Calibration Stategy of CMS Electromagnetic Calorimeter |
title_short | The Calibration Stategy of CMS Electromagnetic Calorimeter |
title_sort | calibration stategy of cms electromagnetic calorimeter |
topic | Detectors and Experimental Techniques |
url | http://cds.cern.ch/record/810667 |
work_keys_str_mv | AT meridianip thecalibrationstategyofcmselectromagneticcalorimeter AT meridianip calibrationstategyofcmselectromagneticcalorimeter |