Cargando…
Response of Production Modules of the ATLAS Tile Calorimeter to Electrons
This paper contains the results of the analyses done on the following electron Tilecal test-beam data: 20 GeV electrons - June 2003 (eta scan, 20 deg scan 90 deg tile scan) and 180 GeV eletrons - August 2003 (eta scan). Correction of the position dependence was applied A Gaussian distribution was fi...
Autor principal: | |
---|---|
Lenguaje: | eng |
Publicado: |
2005
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/906641 |
Sumario: | This paper contains the results of the analyses done on the following electron Tilecal test-beam data: 20 GeV electrons - June 2003 (eta scan, 20 deg scan 90 deg tile scan) and 180 GeV eletrons - August 2003 (eta scan). Correction of the position dependence was applied A Gaussian distribution was fitted to the corrected spectrum and as result the values of the calibration constant and the resolution were obtained. Fit method of the cell signal reconstruction was used. |
---|