Dose equivalent measurements in a strongly pulsed high-energy radiation field
The stray radiation field outside the shielding of high-energy accelerators comprises neutrons, photons and charged particles with a wide range of energies. Often, accelerators operate by accelerating and ejecting short pulses of particles, creating an analogue, pulsed radiation field. The pulses ca...
Autores principales: | , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2004
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1093/rpd/nch177 http://cds.cern.ch/record/909074 |
Sumario: | The stray radiation field outside the shielding of high-energy accelerators comprises neutrons, photons and charged particles with a wide range of energies. Often, accelerators operate by accelerating and ejecting short pulses of particles, creating an analogue, pulsed radiation field. The pulses can be as short as 10 mu s with high instantaneous fluence rates and dose rates. Measurements of average dose equivalent (rate) for radiation protection purposes in these fields present a challenge for instrumentation. The performance of three instruments (i.e. a recombination chamber, the Sievert Instrument and a HANDITEPC) measuring total dose equivalent is compared in a high-energy reference radiation field (CERF) and a strongly pulsed, high-energy radiation field at the CERN proton synchrotron (PS). |
---|