The angular dependence of an Si energy deposition spectrometer response at several radiation sources
An MDU-Liulin spectrometer based on an Si-diode was mainly used during the last few years with the goal to use them for measurements onboard aircraft. To investigate its ability to obtain such measurements, the detector was tested in some radiation reference fields, like /sup 60/Co and other photon...
Autores principales: | , , |
---|---|
Lenguaje: | eng |
Publicado: |
2005
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1016/j.radmeas.2004.06.009 http://cds.cern.ch/record/912961 |
Sumario: | An MDU-Liulin spectrometer based on an Si-diode was mainly used during the last few years with the goal to use them for measurements onboard aircraft. To investigate its ability to obtain such measurements, the detector was tested in some radiation reference fields, like /sup 60/Co and other photon beams, neutrons of an AmBe and /sup 252/Cf sources and in high-energy radiation fields at CERN. Due to the high geometrical asymmetry of the Si-diode semiconductor, an angular dependence of the response would be expected. This work presents analyses and discusses the results of angular dependence studies obtained at the different radiation sources mentioned. It was found that these angular dependences vary with the type and energy of radiation. The influence of these variations on the use as a dosimeter onboard aircraft is also studied and discussed. |
---|