Cargando…
Bulk Etch Rate Measurements and Calibrations of Plastic Nuclear Track Detectors
New calibrations of CR39 and Makrofol nuclear track detectors have been obtained using 158 A GeV Pb (82+) and In (49+) ions/ a new method for the bulk etch rate determination, using both cone height and base diameter measurements was developed. The CR39 charge resolution based on the etch-pit base a...
Autores principales: | , , , , , , , , , , , , , , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2006
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1016/j.nimb.2006.11.056 http://cds.cern.ch/record/992643 |
Sumario: | New calibrations of CR39 and Makrofol nuclear track detectors have been obtained using 158 A GeV Pb (82+) and In (49+) ions/ a new method for the bulk etch rate determination, using both cone height and base diameter measurements was developed. The CR39 charge resolution based on the etch-pit base area measurement is adequate to identify nuclear fragments in the interval 7 <= Z/beta <= 49. For CR39 the detection threshold is at REL~50 MeV cm^2/g, corresponding to a nuclear fragment with Z/beta~7. Base cone area distributions for Makrofol foils exposed to Pb (82+) ions have shown for the first time all peaks due to nuclear fragments with Z > 50/ the distribution of the etched cone heights shows well separated individual peaks for Z/beta = 78 - 83 (charge pickup). The Makrofol detection threshold is at REL 2700 MeV cm^2/g, corresponding to a nuclear fragment with Z/beta~50. |
---|