Cargando…

Performance of the Upgraded LHCf Calorimeter with Gd$_2$SiO$_5$ (GSO) Scintillators Tested by Accelerator Beams

LHCf has reported energy spectra of forward photons and neutral pions at \(\sqrt{s} = 900\) GeV and 7 TeV proton-proton collisions measured at the Large Hadron Collider (LHC). Because the next operation in 2015 is expected under much higher radiation dose, we are upgrading the detectors, especially...

Descripción completa

Detalles Bibliográficos
Autor principal: Makino, Yuya
Lenguaje:eng
Publicado: 2014
Materias:
Acceso en línea:https://dx.doi.org/10.7566/JPSCP.1.013016
http://cds.cern.ch/record/2025920
_version_ 1780947244787171328
author Makino, Yuya
author_facet Makino, Yuya
author_sort Makino, Yuya
collection CERN
description LHCf has reported energy spectra of forward photons and neutral pions at \(\sqrt{s} = 900\) GeV and 7 TeV proton-proton collisions measured at the Large Hadron Collider (LHC). Because the next operation in 2015 is expected under much higher radiation dose, we are upgrading the detectors, especially their scintillators, to be radiation harder. We report the results from the beam tests performed at the Heavy Ion Medical Accelerator in Chiba (HIMAC) in Japan for the basic properties of new sensors. We have confirmed that the new sensors meet our requirements at this stage.
id oai-inspirehep.net-1353203
institution Organización Europea para la Investigación Nuclear
language eng
publishDate 2014
record_format invenio
spelling oai-inspirehep.net-13532032019-09-30T06:29:59Zdoi:10.7566/JPSCP.1.013016http://cds.cern.ch/record/2025920engMakino, YuyaPerformance of the Upgraded LHCf Calorimeter with Gd$_2$SiO$_5$ (GSO) Scintillators Tested by Accelerator BeamsDetectors and Experimental TechniquesLHCf has reported energy spectra of forward photons and neutral pions at \(\sqrt{s} = 900\) GeV and 7 TeV proton-proton collisions measured at the Large Hadron Collider (LHC). Because the next operation in 2015 is expected under much higher radiation dose, we are upgrading the detectors, especially their scintillators, to be radiation harder. We report the results from the beam tests performed at the Heavy Ion Medical Accelerator in Chiba (HIMAC) in Japan for the basic properties of new sensors. We have confirmed that the new sensors meet our requirements at this stage.oai:inspirehep.net:13532032014
spellingShingle Detectors and Experimental Techniques
Makino, Yuya
Performance of the Upgraded LHCf Calorimeter with Gd$_2$SiO$_5$ (GSO) Scintillators Tested by Accelerator Beams
title Performance of the Upgraded LHCf Calorimeter with Gd$_2$SiO$_5$ (GSO) Scintillators Tested by Accelerator Beams
title_full Performance of the Upgraded LHCf Calorimeter with Gd$_2$SiO$_5$ (GSO) Scintillators Tested by Accelerator Beams
title_fullStr Performance of the Upgraded LHCf Calorimeter with Gd$_2$SiO$_5$ (GSO) Scintillators Tested by Accelerator Beams
title_full_unstemmed Performance of the Upgraded LHCf Calorimeter with Gd$_2$SiO$_5$ (GSO) Scintillators Tested by Accelerator Beams
title_short Performance of the Upgraded LHCf Calorimeter with Gd$_2$SiO$_5$ (GSO) Scintillators Tested by Accelerator Beams
title_sort performance of the upgraded lhcf calorimeter with gd$_2$sio$_5$ (gso) scintillators tested by accelerator beams
topic Detectors and Experimental Techniques
url https://dx.doi.org/10.7566/JPSCP.1.013016
http://cds.cern.ch/record/2025920
work_keys_str_mv AT makinoyuya performanceoftheupgradedlhcfcalorimeterwithgd2sio5gsoscintillatorstestedbyacceleratorbeams