ISOLTRAP's multi-reflection time-of-flight mass separator/spectrometer
The online precision mass spectrometer ISOLTRAP at ISOLDE/CERN was recently upgraded by adding a multi-reflection time-of-flight mass separator/spectrometer (MR-ToF MS) between the linear radio-frequency ion trap and the two Penning traps already in place. As a mass separator, the MR-ToF device has...
Autores principales: | , , , , , , , , , , , , , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2013
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1016/j.ijms.2013.03.020 http://cds.cern.ch/record/2279141 |
Sumario: | The online precision mass spectrometer ISOLTRAP at ISOLDE/CERN was recently upgraded by adding a multi-reflection time-of-flight mass separator/spectrometer (MR-ToF MS) between the linear radio-frequency ion trap and the two Penning traps already in place. As a mass separator, the MR-ToF device has improved significantly ISOLTRAP's capability of purification of contaminated ion beams. In addition, the MR-ToF MS can be operated as a mass spectrometer, either to analyze the ISOLDE ion beam or for precision mass measurements of nuclides that are shorter-lived or that have lower yields than those accessible for Penning-trap mass spectrometry. The MR-ToF MS and corresponding components, its integration into ISOLTRAP, and its various operation modes are reviewed. Furthermore, a precision measurement of the 137Eu mass is presented, determined with the help of the MR-ToF device as a mass separator. |
---|