Cargando…

Tuning of the CERN 750 MHz RFQ for medical applications

CERN has built a compact 750 MHz RFQ as an injector for a hadron therapy linac. This RFQ was designed to accelerate protons to an energy of 5 MeV within only 2 m length. It is divided into four segments and equipped with 32 tuners in total. The RFQ length corresponds to 5λ which is considered to be...

Descripción completa

Detalles Bibliográficos
Autores principales: Koubek, Benjamin, Cuvet, Yves, Grudiev, Alexej, Rossi, Carlo, Timmins, Marc
Lenguaje:eng
Publicado: 2017
Materias:
Acceso en línea:https://dx.doi.org/10.18429/JACoW-LINAC2016-THOP09
http://cds.cern.ch/record/2304387
_version_ 1780957462908633088
author Koubek, Benjamin
Cuvet, Yves
Grudiev, Alexej
Rossi, Carlo
Timmins, Marc
author_facet Koubek, Benjamin
Cuvet, Yves
Grudiev, Alexej
Rossi, Carlo
Timmins, Marc
author_sort Koubek, Benjamin
collection CERN
description CERN has built a compact 750 MHz RFQ as an injector for a hadron therapy linac. This RFQ was designed to accelerate protons to an energy of 5 MeV within only 2 m length. It is divided into four segments and equipped with 32 tuners in total. The RFQ length corresponds to 5λ which is considered to be close to the limit for simple field adjustment using tuners. Nevertheless the high frequency results in a sensitive structure and requires careful tuning by means of the alignment of the pumping ports and fixed tuners. This paper gives an overview of the tuning procedure and bead pull measurements of the RFQ.
id oai-inspirehep.net-1633261
institution Organización Europea para la Investigación Nuclear
language eng
publishDate 2017
record_format invenio
spelling oai-inspirehep.net-16332612022-03-07T14:53:01Zdoi:10.18429/JACoW-LINAC2016-THOP09http://cds.cern.ch/record/2304387engKoubek, BenjaminCuvet, YvesGrudiev, AlexejRossi, CarloTimmins, MarcTuning of the CERN 750 MHz RFQ for medical applicationsAccelerators and Storage RingsCERN has built a compact 750 MHz RFQ as an injector for a hadron therapy linac. This RFQ was designed to accelerate protons to an energy of 5 MeV within only 2 m length. It is divided into four segments and equipped with 32 tuners in total. The RFQ length corresponds to 5λ which is considered to be close to the limit for simple field adjustment using tuners. Nevertheless the high frequency results in a sensitive structure and requires careful tuning by means of the alignment of the pumping ports and fixed tuners. This paper gives an overview of the tuning procedure and bead pull measurements of the RFQ.CERN-ACC-2017-349oai:inspirehep.net:16332612017
spellingShingle Accelerators and Storage Rings
Koubek, Benjamin
Cuvet, Yves
Grudiev, Alexej
Rossi, Carlo
Timmins, Marc
Tuning of the CERN 750 MHz RFQ for medical applications
title Tuning of the CERN 750 MHz RFQ for medical applications
title_full Tuning of the CERN 750 MHz RFQ for medical applications
title_fullStr Tuning of the CERN 750 MHz RFQ for medical applications
title_full_unstemmed Tuning of the CERN 750 MHz RFQ for medical applications
title_short Tuning of the CERN 750 MHz RFQ for medical applications
title_sort tuning of the cern 750 mhz rfq for medical applications
topic Accelerators and Storage Rings
url https://dx.doi.org/10.18429/JACoW-LINAC2016-THOP09
http://cds.cern.ch/record/2304387
work_keys_str_mv AT koubekbenjamin tuningofthecern750mhzrfqformedicalapplications
AT cuvetyves tuningofthecern750mhzrfqformedicalapplications
AT grudievalexej tuningofthecern750mhzrfqformedicalapplications
AT rossicarlo tuningofthecern750mhzrfqformedicalapplications
AT timminsmarc tuningofthecern750mhzrfqformedicalapplications