Frequency Scanning Interferometry as New Solution for on-Line Monitoring Inside a Cryostat for the HL-LHC project

For the HL-LHC project, the cryostats of the key components will be equipped permanently with an on-line position monitoring system based on Frequency Scanning Interferometry (FSI). Such a system, based on absolute distance measurement, will determine the position of the inner triplet cold masses w....

Descripción completa

Detalles Bibliográficos
Autores principales: Mainaud Durand, Helene, Dijoud, Thibault, Duquenne, Mathieu, Gayde, Jean-Christophe, Micolon, Frederic, Rude, Vivien, Sosin, Mateusz
Lenguaje:eng
Publicado: 2018
Materias:
Acceso en línea:https://dx.doi.org/10.18429/JACoW-IPAC2018-WEPAF068
http://cds.cern.ch/record/2667534
Descripción
Sumario:For the HL-LHC project, the cryostats of the key components will be equipped permanently with an on-line position monitoring system based on Frequency Scanning Interferometry (FSI). Such a system, based on absolute distance measurement, will determine the position of the inner triplet cold masses w.r.t. their cryostat and the position of the crab cavities also inside their cryostat, within an uncertainty of measurement of 0.1 mm, in a harsh environment: cold temperature of 2 K and high radiation level of the order of 1 MGy. The FSI system was validated first successfully on one LHC dipole cryostat and its associated cold mass to undergo qualification tests under different conditions: warm, vacuum and cold (2K). The FSI system also equips the first crab cavities prototype cryostat. The configuration of the FIS system chosen after simulations, the conditions of tests as well as their results and analysis are presented in this paper.