Sensor array calibration by a single reference in presence of measurand non-linear gradient
A calibration method exploiting a single external reference is proposed for sensors arrays in presence of a measurand non-linear gradient in the system under test. The calibration artifacts due to a non-uniform distribution of the measurand in the system under test, not detected by the single punctu...
Autores principales: | , , , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2018
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1109/I2MTC.2018.8409713 http://cds.cern.ch/record/2682987 |
Sumario: | A calibration method exploiting a single external reference is proposed for sensors arrays in presence of a measurand non-linear gradient in the system under test. The calibration artifacts due to a non-uniform distribution of the measurand in the system under test, not detected by the single punctual reference sensor, are avoided. A proof-of-principle demonstration is carried out within the case study of a thermistors array obtaining compatible result with respect to a standard method which requires a reference sensor for each element of the sensor array. Thus, the complexity (and cost) of standard calbration setups is reduced by using a single reference sensor without sacrifying significantly the calibration accuracy. |
---|