Cargando…
RaDoM: a lung dosimeter for radon progeny
This paper discusses a novel Timepix-based on-line radon dose monitor called RaDoM. The detector provides a direct estimate of the effective dose to the lung via a system of filters having the same absorption characteristics as this organ and a measurement of the PAEC (the potential alpha energy con...
Autores principales: | , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2018
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1088/1748-0221/13/07/P07030 http://cds.cern.ch/record/2665504 |
_version_ | 1780961953533919232 |
---|---|
author | Romano, S Caresana, M Garlati, L Murtas, F Silari, M |
author_facet | Romano, S Caresana, M Garlati, L Murtas, F Silari, M |
author_sort | Romano, S |
collection | CERN |
description | This paper discusses a novel Timepix-based on-line radon dose monitor called RaDoM. The detector provides a direct estimate of the effective dose to the lung via a system of filters having the same absorption characteristics as this organ and a measurement of the PAEC (the potential alpha energy concentration). Measurements performed in a reference radon chamber and in normal environmental conditions showed that RaDoM well follows changes in the radon concentration and is accurate in both long- and short-term measurements. The paper describes the detector and associated software for data acquisition and analysis, and compares the measurement of the lung dose with the value derived from a measurement of the radon concentration. |
id | oai-inspirehep.net-1718422 |
institution | Organización Europea para la Investigación Nuclear |
language | eng |
publishDate | 2018 |
record_format | invenio |
spelling | oai-inspirehep.net-17184222022-08-10T12:29:24Zdoi:10.1088/1748-0221/13/07/P07030http://cds.cern.ch/record/2665504engRomano, SCaresana, MGarlati, LMurtas, FSilari, MRaDoM: a lung dosimeter for radon progenyDetectors and Experimental TechniquesThis paper discusses a novel Timepix-based on-line radon dose monitor called RaDoM. The detector provides a direct estimate of the effective dose to the lung via a system of filters having the same absorption characteristics as this organ and a measurement of the PAEC (the potential alpha energy concentration). Measurements performed in a reference radon chamber and in normal environmental conditions showed that RaDoM well follows changes in the radon concentration and is accurate in both long- and short-term measurements. The paper describes the detector and associated software for data acquisition and analysis, and compares the measurement of the lung dose with the value derived from a measurement of the radon concentration.oai:inspirehep.net:17184222018 |
spellingShingle | Detectors and Experimental Techniques Romano, S Caresana, M Garlati, L Murtas, F Silari, M RaDoM: a lung dosimeter for radon progeny |
title | RaDoM: a lung dosimeter for radon progeny |
title_full | RaDoM: a lung dosimeter for radon progeny |
title_fullStr | RaDoM: a lung dosimeter for radon progeny |
title_full_unstemmed | RaDoM: a lung dosimeter for radon progeny |
title_short | RaDoM: a lung dosimeter for radon progeny |
title_sort | radom: a lung dosimeter for radon progeny |
topic | Detectors and Experimental Techniques |
url | https://dx.doi.org/10.1088/1748-0221/13/07/P07030 http://cds.cern.ch/record/2665504 |
work_keys_str_mv | AT romanos radomalungdosimeterforradonprogeny AT caresanam radomalungdosimeterforradonprogeny AT garlatil radomalungdosimeterforradonprogeny AT murtasf radomalungdosimeterforradonprogeny AT silarim radomalungdosimeterforradonprogeny |