Characteristic study of a quadruple GEM detector in different electric field configurations

A quadruple GEM detector has been assembled and operated using Ar/CO$_2$ 90:10 and 70:30 gas mixtures. Detailed study of the effective gain and energy resolution of the detector are performed using the X-ray spectrum of $^{55}$Fe source. The gain and energy resolution of the detector are measured an...

Descripción completa

Detalles Bibliográficos
Autores principales: Patra, Rajendra Nath, Singaraju, Rama Narayan, Dalal, Somnath, Biswas, Saikat, Viyogi, Yogendra P, Nayak, Tapan K
Lenguaje:eng
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://dx.doi.org/10.1016/j.nima.2018.11.019
http://cds.cern.ch/record/2717431
Descripción
Sumario:A quadruple GEM detector has been assembled and operated using Ar/CO$_2$ 90:10 and 70:30 gas mixtures. Detailed study of the effective gain and energy resolution of the detector are performed using the X-ray spectrum of $^{55}$Fe source. The gain and energy resolution of the detector are measured and compared for two different electric field configurations, the standard field setting and the ion backflow ( IBF ) field setting. It is observed that for the standard field configuration the quadruple GEM detector can operate at lower operating voltages and has better energy resolution compared to IBF field setting.