Cargando…

Single-longitudinal-mode pumped pulsed-dye amplifier for high-resolution laser spectroscopy

The In-Gas-jet Laser Ionization and Spectroscopy (IGLIS) technique relies on narrow-bandwidth, high-peak-power, short-pulse-length (≈10 ns), and high-repetition-rate laser pulses to probe, precisely and efficiently, the hyperfine structure of medium-heavy and heavy isotopes, embedded in a supersonic...

Descripción completa

Detalles Bibliográficos
Autores principales: Verlinde, M, Ferrer, R, Claessens, A, Granados, C A, Kraemer, S, Kudryavtsev, Yu, Li, D, Van den Bergh, P, Van Duppen, P, Verstraelen, E
Lenguaje:eng
Publicado: 2020
Acceso en línea:https://dx.doi.org/10.1063/5.0017985
http://cds.cern.ch/record/2744543
Descripción
Sumario:The In-Gas-jet Laser Ionization and Spectroscopy (IGLIS) technique relies on narrow-bandwidth, high-peak-power, short-pulse-length (≈10 ns), and high-repetition-rate laser pulses to probe, precisely and efficiently, the hyperfine structure of medium-heavy and heavy isotopes, embedded in a supersonic jet. The power and repetition rate requirements of the laser system are met by combining ≈100 W, 8 ns pulse width, 10 kHz commercial Nd:YAG pump lasers with a single-mode continuous wave seeded Pulsed Dye Amplifier (PDA). The common multi-longitudinal-mode operation of these Nd:YAG pump lasers causes, however, undesirable frequency sidebands in the output spectrum of the PDA system, hindering the attainable spectral resolution, a correct interpretation, and an accurate analysis of the hyperfine spectra. In this article, a new prototype Nd:YAG laser is presented, which combined with the PDA system is capable of providing quasi-transform-limited laser pulses at 10 kHz, with only limited losses in laser power. This system reduces any spectral sideband amplitude below a proven upper limit of 0.2% with one order of magnitude extra reduction expected based on simulations. A full characterization of both the Nd:YAG and PDA laser systems is done by studying the temporal and frequency behavior in detail. This study is finalized by a performance benchmark of this combined laser system in the hyperfine spectroscopy of copper isotopes, showcasing its applicability for future IGLIS studies.