Monte Carlo Simulation of the Siemens Biograph Vision PET With Extended Axial Field of View Using Sparse Detector Module Rings Configuration

We report on the NEMA-NU2-2012 performance of a hypothetical Monte Carlo (MC) model, Ex-PET, of the Siemens Biograph Vision positron emission tomography (PET)/CT (Bio-Vis) with sparse detector module rings and extended axial field of view (AFOV). MC simulations were performed with the detector modul...

Descripción completa

Detalles Bibliográficos
Autores principales: Zein, Sara A, Karakatsanis, Nicolas A, Conti, Maurizio, Nehmeh, Sadek A
Lenguaje:eng
Publicado: 2021
Materias:
Acceso en línea:https://dx.doi.org/10.1109/trpms.2020.3034676
http://cds.cern.ch/record/2772590
_version_ 1780971514834714624
author Zein, Sara A
Karakatsanis, Nicolas A
Conti, Maurizio
Nehmeh, Sadek A
author_facet Zein, Sara A
Karakatsanis, Nicolas A
Conti, Maurizio
Nehmeh, Sadek A
author_sort Zein, Sara A
collection CERN
description We report on the NEMA-NU2-2012 performance of a hypothetical Monte Carlo (MC) model, Ex-PET, of the Siemens Biograph Vision positron emission tomography (PET)/CT (Bio-Vis) with sparse detector module rings and extended axial field of view (AFOV). MC simulations were performed with the detector module rings interleaved with 32-mm gaps, equivalent to the axial dimension of each detector module, yielding an AFOV of 48.0 cm (Bio-Vis has 25.6-cm AFOV). 3D-PET acquisition combined with a limited continuous-bed-motion (limited-CBM) was used to compensate for the loss in sensitivity within the gaps’ regions. MC simulations of the Bio-Vis were performed for comparison purposes. All MC simulations were performed using GATE MC toolkit. Ex-PET exhibited 0.49, 0.16, and 0.16 mm deterioration in axial resolution at 1, 10, and 20 cm off-center of the transaxial field of view, respectively, compared to Bio-Vis. Only 1% reduction in system sensitivity and 6% reduction in peak NECR was observed with Ex-PET compared to Bio-Vis. 3D-OSEM image reconstruction, combined with CBM, allowed compensating for the lack of counts within the gaps’ regions. NEMA Image Quality test showed < 6% reduction in contrast recovery with Ex-PET versus Bio-Vis, yet the background variability was increased by up to 8%. The feasibility of PET imaging with an easily adoptable sparse detector configuration was demonstrated. This can lay the pathway for future development of cost-effective PET systems with long and conventional AFOV’s.
id oai-inspirehep.net-1866868
institution Organización Europea para la Investigación Nuclear
language eng
publishDate 2021
record_format invenio
spelling oai-inspirehep.net-18668682022-08-09T14:57:15Zdoi:10.1109/trpms.2020.3034676http://cds.cern.ch/record/2772590engZein, Sara AKarakatsanis, Nicolas AConti, MaurizioNehmeh, Sadek AMonte Carlo Simulation of the Siemens Biograph Vision PET With Extended Axial Field of View Using Sparse Detector Module Rings ConfigurationOtherWe report on the NEMA-NU2-2012 performance of a hypothetical Monte Carlo (MC) model, Ex-PET, of the Siemens Biograph Vision positron emission tomography (PET)/CT (Bio-Vis) with sparse detector module rings and extended axial field of view (AFOV). MC simulations were performed with the detector module rings interleaved with 32-mm gaps, equivalent to the axial dimension of each detector module, yielding an AFOV of 48.0 cm (Bio-Vis has 25.6-cm AFOV). 3D-PET acquisition combined with a limited continuous-bed-motion (limited-CBM) was used to compensate for the loss in sensitivity within the gaps’ regions. MC simulations of the Bio-Vis were performed for comparison purposes. All MC simulations were performed using GATE MC toolkit. Ex-PET exhibited 0.49, 0.16, and 0.16 mm deterioration in axial resolution at 1, 10, and 20 cm off-center of the transaxial field of view, respectively, compared to Bio-Vis. Only 1% reduction in system sensitivity and 6% reduction in peak NECR was observed with Ex-PET compared to Bio-Vis. 3D-OSEM image reconstruction, combined with CBM, allowed compensating for the lack of counts within the gaps’ regions. NEMA Image Quality test showed < 6% reduction in contrast recovery with Ex-PET versus Bio-Vis, yet the background variability was increased by up to 8%. The feasibility of PET imaging with an easily adoptable sparse detector configuration was demonstrated. This can lay the pathway for future development of cost-effective PET systems with long and conventional AFOV’s.oai:inspirehep.net:18668682021
spellingShingle Other
Zein, Sara A
Karakatsanis, Nicolas A
Conti, Maurizio
Nehmeh, Sadek A
Monte Carlo Simulation of the Siemens Biograph Vision PET With Extended Axial Field of View Using Sparse Detector Module Rings Configuration
title Monte Carlo Simulation of the Siemens Biograph Vision PET With Extended Axial Field of View Using Sparse Detector Module Rings Configuration
title_full Monte Carlo Simulation of the Siemens Biograph Vision PET With Extended Axial Field of View Using Sparse Detector Module Rings Configuration
title_fullStr Monte Carlo Simulation of the Siemens Biograph Vision PET With Extended Axial Field of View Using Sparse Detector Module Rings Configuration
title_full_unstemmed Monte Carlo Simulation of the Siemens Biograph Vision PET With Extended Axial Field of View Using Sparse Detector Module Rings Configuration
title_short Monte Carlo Simulation of the Siemens Biograph Vision PET With Extended Axial Field of View Using Sparse Detector Module Rings Configuration
title_sort monte carlo simulation of the siemens biograph vision pet with extended axial field of view using sparse detector module rings configuration
topic Other
url https://dx.doi.org/10.1109/trpms.2020.3034676
http://cds.cern.ch/record/2772590
work_keys_str_mv AT zeinsaraa montecarlosimulationofthesiemensbiographvisionpetwithextendedaxialfieldofviewusingsparsedetectormoduleringsconfiguration
AT karakatsanisnicolasa montecarlosimulationofthesiemensbiographvisionpetwithextendedaxialfieldofviewusingsparsedetectormoduleringsconfiguration
AT contimaurizio montecarlosimulationofthesiemensbiographvisionpetwithextendedaxialfieldofviewusingsparsedetectormoduleringsconfiguration
AT nehmehsadeka montecarlosimulationofthesiemensbiographvisionpetwithextendedaxialfieldofviewusingsparsedetectormoduleringsconfiguration