Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function

Background: The ABL90 FLEX PLUS (Radiometer) is a blood gas analyzer that also provides creatinine (Cr) and blood urea nitrogen (BUN) results. We assessed the accuracy of the ABL90 FLEX PLUS to measure Cr and BUN and find suitable candidate specimens against primary specimens (heparinized whole-bloo...

Descripción completa

Detalles Bibliográficos
Autores principales: Lim, Ha-Jin, Lee, Seung-Yeob, Choi, Hyun-Jung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10003958/
https://www.ncbi.nlm.nih.gov/pubmed/36902727
http://dx.doi.org/10.3390/jcm12051940
_version_ 1784904724932198400
author Lim, Ha-Jin
Lee, Seung-Yeob
Choi, Hyun-Jung
author_facet Lim, Ha-Jin
Lee, Seung-Yeob
Choi, Hyun-Jung
author_sort Lim, Ha-Jin
collection PubMed
description Background: The ABL90 FLEX PLUS (Radiometer) is a blood gas analyzer that also provides creatinine (Cr) and blood urea nitrogen (BUN) results. We assessed the accuracy of the ABL90 FLEX PLUS to measure Cr and BUN and find suitable candidate specimens against primary specimens (heparinized whole-blood (H-WB)). Methods: Paired H-WB, serum, and sodium-citrated whole-blood (C-WB) samples (105) were collected. The Cr and BUN levels in the H-WB using the ABL90 FLEX PLUS were compared with those of the serum using four automated chemistry analyzers. The suitability of the candidate specimens was assessed at each medical decision level according to the CLSI guideline EP35-ED1. Results: The respective mean differences of the ABL90 FLEX PLUS for the Cr and BUN were below −0.10 and −3.51 mg/dL compared to the other analyzers. The systematic differences between the serum and the H-WB at the low, medium, and high medical decision levels were all 0% for Cr, but those of the C-WB were −12.96%, −11.81%, and −11.30%, respectively. Regarding imprecision, the SD(serum)/SD(H-WB) ratios at each level were 0.14, 1.41, and 0.68, whereas the SD(C-WB)/SD(H-WB) ratios were 0.35, 2.00, and 0.73, respectively. Conclusions: The ABL90 FLEX PLUS provided Cr and BUN results comparable with the four widely used analyzers. Among the candidates, the serum was suitable for Cr testing using the ABL90 FLEX PLUS, while the C-WB did not satisfy the acceptance criteria.
format Online
Article
Text
id pubmed-10003958
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-100039582023-03-11 Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function Lim, Ha-Jin Lee, Seung-Yeob Choi, Hyun-Jung J Clin Med Article Background: The ABL90 FLEX PLUS (Radiometer) is a blood gas analyzer that also provides creatinine (Cr) and blood urea nitrogen (BUN) results. We assessed the accuracy of the ABL90 FLEX PLUS to measure Cr and BUN and find suitable candidate specimens against primary specimens (heparinized whole-blood (H-WB)). Methods: Paired H-WB, serum, and sodium-citrated whole-blood (C-WB) samples (105) were collected. The Cr and BUN levels in the H-WB using the ABL90 FLEX PLUS were compared with those of the serum using four automated chemistry analyzers. The suitability of the candidate specimens was assessed at each medical decision level according to the CLSI guideline EP35-ED1. Results: The respective mean differences of the ABL90 FLEX PLUS for the Cr and BUN were below −0.10 and −3.51 mg/dL compared to the other analyzers. The systematic differences between the serum and the H-WB at the low, medium, and high medical decision levels were all 0% for Cr, but those of the C-WB were −12.96%, −11.81%, and −11.30%, respectively. Regarding imprecision, the SD(serum)/SD(H-WB) ratios at each level were 0.14, 1.41, and 0.68, whereas the SD(C-WB)/SD(H-WB) ratios were 0.35, 2.00, and 0.73, respectively. Conclusions: The ABL90 FLEX PLUS provided Cr and BUN results comparable with the four widely used analyzers. Among the candidates, the serum was suitable for Cr testing using the ABL90 FLEX PLUS, while the C-WB did not satisfy the acceptance criteria. MDPI 2023-03-01 /pmc/articles/PMC10003958/ /pubmed/36902727 http://dx.doi.org/10.3390/jcm12051940 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Lim, Ha-Jin
Lee, Seung-Yeob
Choi, Hyun-Jung
Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function
title Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function
title_full Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function
title_fullStr Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function
title_full_unstemmed Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function
title_short Evaluation of the Accuracy of Cr and BUN Using the ABL90 FLEX PLUS Blood Gas Analyzer and the Equivalence of Candidate Specimens for Assessment of Renal Function
title_sort evaluation of the accuracy of cr and bun using the abl90 flex plus blood gas analyzer and the equivalence of candidate specimens for assessment of renal function
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10003958/
https://www.ncbi.nlm.nih.gov/pubmed/36902727
http://dx.doi.org/10.3390/jcm12051940
work_keys_str_mv AT limhajin evaluationoftheaccuracyofcrandbunusingtheabl90flexplusbloodgasanalyzerandtheequivalenceofcandidatespecimensforassessmentofrenalfunction
AT leeseungyeob evaluationoftheaccuracyofcrandbunusingtheabl90flexplusbloodgasanalyzerandtheequivalenceofcandidatespecimensforassessmentofrenalfunction
AT choihyunjung evaluationoftheaccuracyofcrandbunusingtheabl90flexplusbloodgasanalyzerandtheequivalenceofcandidatespecimensforassessmentofrenalfunction