Radiation dose reduction using deep learning-based image reconstruction for a low-dose chest computed tomography protocol: a phantom study

BACKGROUND: The aim of this study was to compare the dose reduction potential and image quality of deep learning-based image reconstruction (DLIR) with those of filtered back-projection (FBP) and iterative reconstruction (IR) and to determine the clinically usable dose of DLIR for low-dose chest com...

Descripción completa

Detalles Bibliográficos
Autores principales: Jung, Yunsub, Hur, Jin, Han, Kyunghwa, Imai, Yasuhiro, Hong, Yoo Jin, Im, Dong Jin, Lee, Kye Ho, Desnoyers, Melissa, Thomsen, Brian, Shigemasa, Risa, Um, Kyounga, Jang, Kyungeun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: AME Publishing Company 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10006148/
https://www.ncbi.nlm.nih.gov/pubmed/36915339
http://dx.doi.org/10.21037/qims-22-618
Descripción
Sumario:BACKGROUND: The aim of this study was to compare the dose reduction potential and image quality of deep learning-based image reconstruction (DLIR) with those of filtered back-projection (FBP) and iterative reconstruction (IR) and to determine the clinically usable dose of DLIR for low-dose chest computed tomography (LDCT) scans. METHODS: Multi-slice computed tomography (CT) scans of a chest phantom were performed with various tube voltages and tube currents, and the images were reconstructed using seven methods to control the amount of noise reduction: FBP, three stages of IR, and three stages of DLIR. For subjective image analysis, four radiologists compared 48 image data sets with reference images and rated on a 5-point scale. For quantitative image analysis, the signal to noise ratio (SNR), contrast to noise ratio (CNR), nodule volume, and nodule diameter were measured. RESULTS: In the subjective analysis, DLIR-Low (0.46 mGy), DLIR-Medium (0.31 mGy), and DLIR-High (0.18 mGy) images showed similar quality to the FBP (2.47 mGy) image. Under the same dose conditions, the SNR and CNR were higher with DLIR-High than with FBP and all the IR methods (all P<0.05). The nodule volume and size with DLIR-High were significantly closer to the real volume than with FBP and all the IR methods (all P<0.001). CONCLUSIONS: DLIR can improve the image quality of LDCT compared to FBP and IR. In addition, the appropriate effective dose for LDCT would be 0.24 mGy with DLIR-High.