Cargando…

Defibrillation effectiveness and safety of the shock waveform used in a contemporary wearable cardioverter defibrillator: Results from animal and human studies

INTRODUCTION: The wearable cardioverter defibrillator (WCD) is used to protect patients at risk for sudden cardiac arrest. We examined defibrillation efficacy and safety of a biphasic truncated exponential waveform designed for use in a contemporary WCD in three animal studies and a human study. MET...

Descripción completa

Detalles Bibliográficos
Autores principales: Gleva, Marye J., Sullivan, Joseph, Crawford, Thomas C., Walcott, Greg, Birgersdotter-Green, Ulrika, Branch, Kelley R., Doshi, Rahul N., Kivilaid, Kaisa, Brennan, Kelly, Rowbotham, Ron K., Gustavson, Laura M., Poole, Jeanne E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10013906/
https://www.ncbi.nlm.nih.gov/pubmed/36917566
http://dx.doi.org/10.1371/journal.pone.0281340
Descripción
Sumario:INTRODUCTION: The wearable cardioverter defibrillator (WCD) is used to protect patients at risk for sudden cardiac arrest. We examined defibrillation efficacy and safety of a biphasic truncated exponential waveform designed for use in a contemporary WCD in three animal studies and a human study. METHODS: Animal (swine) studies: #1: Efficacy comparison of a 170J BTE waveform (SHOCK A) to a 150J BTE waveform (SHOCK B) that approximates another commercially available waveform. Primary endpoint first shock success rate. #2: Efficacy comparison of the two waveforms at attenuated charge voltages in swine at three prespecified impedances. Primary endpoint first shock success rate. #3: Safety comparison of SHOCK A and SHOCK B in swine. Primary endpoint cardiac biomarker level changes baseline to 6 and 24 hours post-shock. Human Study: Efficacy comparison of SHOCK A to prespecified goal and safety evaluation. Primary endpoint cumulative first and second shock success rate. Safety endpoint adverse events. RESULTS: Animal Studies #1: 120 VF episodes in six swine. First shock success rates for SHOCK A and SHOCK B were 100%; SHOCK A non-inferior to SHOCK B (entire 95% CI of rate difference above -10% margin, p < .001). #2: 2,160 VF episodes in thirty-six swine. Attenuated SHOCK A was non-inferior to attenuated SHOCK B at each impedance (entire 95% CI of rate difference above -10% margin, p < .001). #3: Ten swine, five shocked five times each with SHOCK A, five shocked five times each with SHOCK B. No significant difference in troponin I (p = 0.658) or creatine phosphokinase (p = 0.855) changes from baseline between SHOCK A and SHOCK B. Human Study: Thirteen patients, 100% VF conversion rate. Mild skin irritation from adhesive defibrillation pads in three patients. CONCLUSIONS: The BTE waveform effectively and safely terminated induced VF in swine and a small sample in humans. TRIAL REGISTRATION: Human study clinical trial registration: URL: https://clinicaltrials.gov; Unique identifier: NCT04132466.