Gas Dynamic Virtual Nozzle Sprayer for an Introduction of Liquid Samples in Atmospheric Pressure Ionization Mass Spectrometry

[Image: see text] Electrospray may exhibit inadequate ionization efficiency in some applications. In such cases, atmospheric-pressure chemical ionization (APCI) and photoionization (APPI) can be used. Despite a wide application potential, no APCI and APPI sources dedicated to very low sample flow ra...

Descripción completa

Detalles Bibliográficos
Autores principales: Kloudová, Barbora, Strmeň, Timotej, Vrkoslav, Vladimír, Chára, Zdeněk, Pačes, Ondřej, Cvačka, Josef
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2023
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10016749/
https://www.ncbi.nlm.nih.gov/pubmed/36800482
http://dx.doi.org/10.1021/acs.analchem.2c05349
Descripción
Sumario:[Image: see text] Electrospray may exhibit inadequate ionization efficiency in some applications. In such cases, atmospheric-pressure chemical ionization (APCI) and photoionization (APPI) can be used. Despite a wide application potential, no APCI and APPI sources dedicated to very low sample flow rates exist on the market. Since the ion source performance depends on the transfer of analytes from the liquid to the gas phase, a nebulizer is a critical component of an ion source. Here, we report on the nebulizer with a gas dynamic virtual nozzle (GDVN) and its applicability in APCI at microliter-per-minute flow rates. Nebulizers differing by geometrical parameters were fabricated and characterized regarding the jet breakup regime, droplet size, droplet velocity, and spray angle for liquid flow rates of 0.75–15.0 μL/min. A micro-APCI source with the GDVN nebulizer behaved as a mass-flow-sensitive detector and provided stable and intense analyte signals. Compared to a classical APCI source, an order of magnitude lower detection limit for verapamil was achieved. Mass spectra recorded with the nebulizer in dripping and jetting modes were almost identical and did not differ from normal APCI spectra. Clogging never occurred during the experiments, indicating the high robustness of the nebulizer. Low-flow-rate APCI and APPI sources with a GDVN sprayer promise new applications for low- and medium-polar analytes.