Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source dual-energy CT

OBJECTIVE: Subjective and objective image quality comparison of bone microstructure and disease-related abnormalities in multiple myeloma patients using a 1st-generation dual-source photon-counting detector CT(DS-PCD-CT) and a 2nd-generation dual-source dual-energy (energy-integrating detector) CT (...

Descripción completa

Detalles Bibliográficos
Autores principales: Winkelmann, Moritz T., Hagen, Florian, Le-Yannou, Lucie, Weiss, Jakob, Riffel, Philipp, Gutjahr, Ralf, Faby, Sebastian, Nikolaou, Konstantin, Horger, Marius
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10017628/
https://www.ncbi.nlm.nih.gov/pubmed/36350390
http://dx.doi.org/10.1007/s00330-022-09225-0
_version_ 1784907631753691136
author Winkelmann, Moritz T.
Hagen, Florian
Le-Yannou, Lucie
Weiss, Jakob
Riffel, Philipp
Gutjahr, Ralf
Faby, Sebastian
Nikolaou, Konstantin
Horger, Marius
author_facet Winkelmann, Moritz T.
Hagen, Florian
Le-Yannou, Lucie
Weiss, Jakob
Riffel, Philipp
Gutjahr, Ralf
Faby, Sebastian
Nikolaou, Konstantin
Horger, Marius
author_sort Winkelmann, Moritz T.
collection PubMed
description OBJECTIVE: Subjective and objective image quality comparison of bone microstructure and disease-related abnormalities in multiple myeloma patients using a 1st-generation dual-source photon-counting detector CT(DS-PCD-CT) and a 2nd-generation dual-source dual-energy (energy-integrating detector) CT (DS-EID-CT). METHODS: Fifty multiple myeloma patients (mean age 67.7 ± 10.9 years,16 females) were prospectively enrolled. Unenhanced whole-body CTs were clinically indicated and performed on DS-EID-CT and DS-PCD-CT (median time difference: 12 months). DS-PCD-CT was performed in Quantumplus UHR mode and DS-EID-CT was performed using dual-energy mode. DS-PCD-CT kernel was set at Br64 with Quantum iterative reconstruction strength Q1; for DS-EID-CT a comparable I70f kernel with SAFIRE iterative reconstruction strength 1 was used. Two independent radiologists assessed image quality subjectively using a 5-point Likert scale considering delineation and sharpness of trabecular bone and lytic bone lesions in the spine and pelvic bones. Additionally, ImageJ was used for quantification of bony septa inside the cancellous bone and through or the edges of osteolysis. RESULTS: Overall quality as well as detectability and sharpness in the delineation of lytic bone lesions were superior for DS-PCD-CT compared with DS-EID-CT (p < 0.0001). The inter-reader agreement for subjective image quality readings showed excellent consistency(α = 94.2–98.8). CTDI and DLP mean values for DS-PCD-CT and DS-EID-CT were 1107.4 ± 247.6 mGy*cm and 8.2 ± 1.8 mGy vs. 1344.3 ± 204.6 mGy*cm and 10.1 ± 1.9 mGy. The quantitative metric for bone microstructure in the femoral head showed significantly better visualization of trabeculae in DS-PCD-CT compared with DS-EID-CT (p < 0.0001). Quantitative analyses of edge sharpness of osteolysis showed significant steeper edges for DS-PCD-CT (p < 0.0001). CONCLUSION: DS-PCD-CT significantly improves spatial resolution of bony microstructure and lytic bone lesions compared to DS-EID-CT. KEY POINTS: • Application of photon-counting detector CT is superior to dual-source dual-energy integrating detector in clinical workup of multiple myeloma patients. • Compared to energy integrating detectors, photon-counting detectors significantly increase the spatial resolution of bone microstructure including disease-related lytic bone lesions in patients with multiple myeloma.
format Online
Article
Text
id pubmed-10017628
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-100176282023-03-17 Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source dual-energy CT Winkelmann, Moritz T. Hagen, Florian Le-Yannou, Lucie Weiss, Jakob Riffel, Philipp Gutjahr, Ralf Faby, Sebastian Nikolaou, Konstantin Horger, Marius Eur Radiol Computed Tomography OBJECTIVE: Subjective and objective image quality comparison of bone microstructure and disease-related abnormalities in multiple myeloma patients using a 1st-generation dual-source photon-counting detector CT(DS-PCD-CT) and a 2nd-generation dual-source dual-energy (energy-integrating detector) CT (DS-EID-CT). METHODS: Fifty multiple myeloma patients (mean age 67.7 ± 10.9 years,16 females) were prospectively enrolled. Unenhanced whole-body CTs were clinically indicated and performed on DS-EID-CT and DS-PCD-CT (median time difference: 12 months). DS-PCD-CT was performed in Quantumplus UHR mode and DS-EID-CT was performed using dual-energy mode. DS-PCD-CT kernel was set at Br64 with Quantum iterative reconstruction strength Q1; for DS-EID-CT a comparable I70f kernel with SAFIRE iterative reconstruction strength 1 was used. Two independent radiologists assessed image quality subjectively using a 5-point Likert scale considering delineation and sharpness of trabecular bone and lytic bone lesions in the spine and pelvic bones. Additionally, ImageJ was used for quantification of bony septa inside the cancellous bone and through or the edges of osteolysis. RESULTS: Overall quality as well as detectability and sharpness in the delineation of lytic bone lesions were superior for DS-PCD-CT compared with DS-EID-CT (p < 0.0001). The inter-reader agreement for subjective image quality readings showed excellent consistency(α = 94.2–98.8). CTDI and DLP mean values for DS-PCD-CT and DS-EID-CT were 1107.4 ± 247.6 mGy*cm and 8.2 ± 1.8 mGy vs. 1344.3 ± 204.6 mGy*cm and 10.1 ± 1.9 mGy. The quantitative metric for bone microstructure in the femoral head showed significantly better visualization of trabeculae in DS-PCD-CT compared with DS-EID-CT (p < 0.0001). Quantitative analyses of edge sharpness of osteolysis showed significant steeper edges for DS-PCD-CT (p < 0.0001). CONCLUSION: DS-PCD-CT significantly improves spatial resolution of bony microstructure and lytic bone lesions compared to DS-EID-CT. KEY POINTS: • Application of photon-counting detector CT is superior to dual-source dual-energy integrating detector in clinical workup of multiple myeloma patients. • Compared to energy integrating detectors, photon-counting detectors significantly increase the spatial resolution of bone microstructure including disease-related lytic bone lesions in patients with multiple myeloma. Springer Berlin Heidelberg 2022-11-09 2023 /pmc/articles/PMC10017628/ /pubmed/36350390 http://dx.doi.org/10.1007/s00330-022-09225-0 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Computed Tomography
Winkelmann, Moritz T.
Hagen, Florian
Le-Yannou, Lucie
Weiss, Jakob
Riffel, Philipp
Gutjahr, Ralf
Faby, Sebastian
Nikolaou, Konstantin
Horger, Marius
Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source dual-energy CT
title Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source dual-energy CT
title_full Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source dual-energy CT
title_fullStr Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source dual-energy CT
title_full_unstemmed Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source dual-energy CT
title_short Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source dual-energy CT
title_sort myeloma bone disease imaging on a 1st-generation clinical photon-counting detector ct vs. 2nd-generation dual-source dual-energy ct
topic Computed Tomography
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10017628/
https://www.ncbi.nlm.nih.gov/pubmed/36350390
http://dx.doi.org/10.1007/s00330-022-09225-0
work_keys_str_mv AT winkelmannmoritzt myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct
AT hagenflorian myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct
AT leyannoulucie myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct
AT weissjakob myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct
AT riffelphilipp myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct
AT gutjahrralf myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct
AT fabysebastian myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct
AT nikolaoukonstantin myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct
AT horgermarius myelomabonediseaseimagingona1stgenerationclinicalphotoncountingdetectorctvs2ndgenerationdualsourcedualenergyct