Cargando…

The Impact of Data Management on the Achievable Dose and Efficiency of Mammography and Radiography During the COVID-19 Era: A Facility-Based Cohort Study

PURPOSE: To evaluate the impact of using computational data management resources and analytical software on radiation doses in mammography and radiography during the COVID-19 pandemic, develop departmental diagnostic reference levels (DRLs), and describe achievable doses (ADs) for mammography and ra...

Descripción completa

Detalles Bibliográficos
Autores principales: Hegazi, Tarek Mohammed, AlSharydah, Abdulaziz Mohammad, Alfawaz, Iba, Al-Muhanna, Afnan Fahad, Faisal, Sarah Yousef
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10024472/
https://www.ncbi.nlm.nih.gov/pubmed/36941927
http://dx.doi.org/10.2147/RMHP.S389960
Descripción
Sumario:PURPOSE: To evaluate the impact of using computational data management resources and analytical software on radiation doses in mammography and radiography during the COVID-19 pandemic, develop departmental diagnostic reference levels (DRLs), and describe achievable doses (ADs) for mammography and radiography based on measured dose parameters. PATIENTS AND METHODS: This ambispective cohort study enrolled 795 and 12,115 patients who underwent mammography and radiography, respectively, at the King Fahd Hospital of the University, Al-Khobar City, Saudi Arabia between May 25 and November 4, 2021. Demographic data were acquired from patients’ electronic medical charts. Data on mammographic and radiographic dose determinants were acquired from the data management software. Based on the time when the data management software was operational in the institute, the study was divided into the pre-implementation and post-implementation phases. Continuous and categorical variables were compared between the two phases using an unpaired t-test and the chi-square test. RESULTS: The median accumulated average glandular dose (AGD; a mammographic dose determinant) in the post-implementation phase was three-fold higher than that in the pre-implementation phase. The average mammographic exposure time in the post-implementation phase was 16.3 ms shorter than that in the pre-implementation phase. Furthermore, the median values of the dose area product ([DAP], a radiographic dose determinant) were 9.72 and 19.4 cGycm(2) in the pre-implementation and post-implementation phases, respectively. CONCLUSION: Although the data management software used in this study helped reduce the radiation exposure time by 16.3 ms in mammography, its impact on the mean accumulated AGD was unfavorable. Similarly, radiographic exposure indices, including DAP, tube voltage, tube current, and exposure time, were not significantly different after the data management software was implemented. Close monitoring of patient radiation doses in mammography and radiography, and dose reduction will become possible if imaging facilities use DRLs and ADs via automated systems.