Factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmH(2)O during mechanical ventilation: A cross-sectional study

OBJECTIVES: The cuff pressures > 30 cmH(2)O may create a seal in the trachea. The objective of this study was to identify risk factors associated with lack of tracheal sealing by an endotracheal cuff inflated to > 30 cmH(2)O in patients undergoing mechanical ventilation. METHODS: This prospect...

Descripción completa

Detalles Bibliográficos
Autores principales: Wu, Hong-Lei, Shi, Hai-Yan, Shi, Jia-hai, Shen, Wang-Qin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Professional Medical Publications 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10025741/
https://www.ncbi.nlm.nih.gov/pubmed/36950396
http://dx.doi.org/10.12669/pjms.39.2.5672
_version_ 1784909403000930304
author Wu, Hong-Lei
Shi, Hai-Yan
Shi, Jia-hai
Shen, Wang-Qin
author_facet Wu, Hong-Lei
Shi, Hai-Yan
Shi, Jia-hai
Shen, Wang-Qin
author_sort Wu, Hong-Lei
collection PubMed
description OBJECTIVES: The cuff pressures > 30 cmH(2)O may create a seal in the trachea. The objective of this study was to identify risk factors associated with lack of tracheal sealing by an endotracheal cuff inflated to > 30 cmH(2)O in patients undergoing mechanical ventilation. METHODS: This prospective cross-sectional study was conducted from 2019 to 2020 in the cardiothoracic intensive care unit and respiratory medical care unit of a Hospital in Nantong, China. Patients aged >16 years undergoing cardiothoracic surgery with mechanical ventilation using endotracheal intubation were included. Patient characteristics and ventilator parameters were analyzed. Cuff pressure was maintained with the minimum leak technique (MLT) and measured with a cuff pressure gauge. Cuff pressure was measured for 30 seconds when ventilation was accompanied by no leak, simultaneously detected by the ventilator or auscultation with a stethoscope. RESULT: Of 352 patients undergoing mechanical ventilation, 51 patients (14.5%) had a cuff pressure of >30 cmH(2)O. Multivariable analysis showed that cuff manufacturer (Guangzhou Weili) and nasal endotracheal intubation significantly increased the risk of an unsealed trachea. Peak inspiratory pressure, cuff diameter and male sex had a strong inverse association with an unsealed trachea. CONCLUSIONS: These findings suggest that an endotracheal cuff pressure of 20 to 30 cmH(2)O is adequate for most patients, but lack of a tracheal seal still occurs in a small number of people. An unsealed trachea is most likely because cuff and tracheal diameters do not match. Clinical Trial Registration: http://www.chictr.org.cn/index.aspx Unique identifier: ChiCTR-COC-15006459.
format Online
Article
Text
id pubmed-10025741
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Professional Medical Publications
record_format MEDLINE/PubMed
spelling pubmed-100257412023-03-21 Factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmH(2)O during mechanical ventilation: A cross-sectional study Wu, Hong-Lei Shi, Hai-Yan Shi, Jia-hai Shen, Wang-Qin Pak J Med Sci Original Article OBJECTIVES: The cuff pressures > 30 cmH(2)O may create a seal in the trachea. The objective of this study was to identify risk factors associated with lack of tracheal sealing by an endotracheal cuff inflated to > 30 cmH(2)O in patients undergoing mechanical ventilation. METHODS: This prospective cross-sectional study was conducted from 2019 to 2020 in the cardiothoracic intensive care unit and respiratory medical care unit of a Hospital in Nantong, China. Patients aged >16 years undergoing cardiothoracic surgery with mechanical ventilation using endotracheal intubation were included. Patient characteristics and ventilator parameters were analyzed. Cuff pressure was maintained with the minimum leak technique (MLT) and measured with a cuff pressure gauge. Cuff pressure was measured for 30 seconds when ventilation was accompanied by no leak, simultaneously detected by the ventilator or auscultation with a stethoscope. RESULT: Of 352 patients undergoing mechanical ventilation, 51 patients (14.5%) had a cuff pressure of >30 cmH(2)O. Multivariable analysis showed that cuff manufacturer (Guangzhou Weili) and nasal endotracheal intubation significantly increased the risk of an unsealed trachea. Peak inspiratory pressure, cuff diameter and male sex had a strong inverse association with an unsealed trachea. CONCLUSIONS: These findings suggest that an endotracheal cuff pressure of 20 to 30 cmH(2)O is adequate for most patients, but lack of a tracheal seal still occurs in a small number of people. An unsealed trachea is most likely because cuff and tracheal diameters do not match. Clinical Trial Registration: http://www.chictr.org.cn/index.aspx Unique identifier: ChiCTR-COC-15006459. Professional Medical Publications 2023 /pmc/articles/PMC10025741/ /pubmed/36950396 http://dx.doi.org/10.12669/pjms.39.2.5672 Text en Copyright: © Pakistan Journal of Medical Sciences https://creativecommons.org/licenses/by/3.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0 (https://creativecommons.org/licenses/by/3.0/) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Wu, Hong-Lei
Shi, Hai-Yan
Shi, Jia-hai
Shen, Wang-Qin
Factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmH(2)O during mechanical ventilation: A cross-sectional study
title Factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmH(2)O during mechanical ventilation: A cross-sectional study
title_full Factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmH(2)O during mechanical ventilation: A cross-sectional study
title_fullStr Factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmH(2)O during mechanical ventilation: A cross-sectional study
title_full_unstemmed Factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmH(2)O during mechanical ventilation: A cross-sectional study
title_short Factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmH(2)O during mechanical ventilation: A cross-sectional study
title_sort factors associated with lack of tracheal sealing by a cuff inflated to more than 30 cmh(2)o during mechanical ventilation: a cross-sectional study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10025741/
https://www.ncbi.nlm.nih.gov/pubmed/36950396
http://dx.doi.org/10.12669/pjms.39.2.5672
work_keys_str_mv AT wuhonglei factorsassociatedwithlackoftrachealsealingbyacuffinflatedtomorethan30cmh2oduringmechanicalventilationacrosssectionalstudy
AT shihaiyan factorsassociatedwithlackoftrachealsealingbyacuffinflatedtomorethan30cmh2oduringmechanicalventilationacrosssectionalstudy
AT shijiahai factorsassociatedwithlackoftrachealsealingbyacuffinflatedtomorethan30cmh2oduringmechanicalventilationacrosssectionalstudy
AT shenwangqin factorsassociatedwithlackoftrachealsealingbyacuffinflatedtomorethan30cmh2oduringmechanicalventilationacrosssectionalstudy