Neuronavigation maximizes accuracy and precision in TMS positioning: Evidence from 11,230 distance, angle, and electric field modeling measurements

BACKGROUND: Researchers and clinicians have traditionally relied on elastic caps with markings to reposition the transcranial magnetic stimulation (TMS) coil between trains and sessions. Newer neuronavigation technology co-registers the patient’s head and structural magnetic resonance imaging (MRI)...

Descripción completa

Detalles Bibliográficos
Autores principales: Caulfield, Kevin A., Fleischmann, Holly H., Cox, Claire E., Wolf, Julia P., George, Mark S., McTeague, Lisa M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10026380/
https://www.ncbi.nlm.nih.gov/pubmed/36031059
http://dx.doi.org/10.1016/j.brs.2022.08.013
_version_ 1784909529596559360
author Caulfield, Kevin A.
Fleischmann, Holly H.
Cox, Claire E.
Wolf, Julia P.
George, Mark S.
McTeague, Lisa M.
author_facet Caulfield, Kevin A.
Fleischmann, Holly H.
Cox, Claire E.
Wolf, Julia P.
George, Mark S.
McTeague, Lisa M.
author_sort Caulfield, Kevin A.
collection PubMed
description BACKGROUND: Researchers and clinicians have traditionally relied on elastic caps with markings to reposition the transcranial magnetic stimulation (TMS) coil between trains and sessions. Newer neuronavigation technology co-registers the patient’s head and structural magnetic resonance imaging (MRI) scan, providing the researcher with real-time feedback about how to adjust the coil to be on-target. However, there has been no head to head comparison of accuracy and precision across treatment sessions. OBJECTIVE: /Hypothesis: In this two-part study, we compared elastic cap and neuronavigation targeting methodologies on distance, angle, and electric field (E-field) magnitude values. METHODS: In 42 participants receiving up to 50 total accelerated rTMS sessions in 5 days, we compared cap and neuronavigation targeting approaches in 3408 distance and 6816 angle measurements. In Experiment 1, TMS administrators saved an on-target neuronavigation location at Beam F3, which served as the landmark for all other measurements. Next, the operators placed the TMS coil based on cap markings or neuronavigation software to measure the distance and angle differences from the on-target sample. In Experiment 2, we saved each XYZ coordinate of the TMS coil from cap and neuronavigation targeting in 12 participants to compare the E-field magnitude differences at the cortical prefrontal target in 1106 cap and neuronavigation models. RESULTS: Cap targeting was significantly off-target for distance, placing the coil an average of 10.66 mm off-target (Standard error of the mean; SEM = 0.19 mm) compared to 0.3 mm (SEM = 0.03 mm) for neuronavigation (p < 0.0001). Cap targeting also significantly deviated for angles off-target, averaging 7.79 roll/pitch degrees (SEM = 1.07°) off-target and 5.99 yaw degrees (SEM = 0.12°) off-target; in comparison, neuronavigation targeting positioned the coil 0.34 roll/pitch degrees (SEM = 0.01°) and 0.22 yaw (SEM = 0.004°) off-target (both p < 0.0001). Further analyses revealed that there were significant inter-operator differences on distance and angle positioning for F3 (all p < 0.05), but not neuronavigation. Lastly, cap targeting resulted in significantly lower E-fields at the intended prefrontal cortical target, with equivalent E-fields as 110.7% motor threshold (MT; range = 58.3–127.4%) stimulation vs. 119.9% MT (range = 115–123.3%) from neuronavigated targeting with 120% MT stimulation applied (p < 0.001). CONCLUSIONS: Cap-based targeting is an inherent source of target variability compared to neuronavigation. Additionally, cap-based coil placement is more prone to differences across operators. Off-target coil placement secondary to cap-based measurements results in significantly lower amounts of stimulation reaching the cortical target, with some individuals receiving only 48.6% of the intended on-target E-field. Neuronavigation technology enables more precise and accurate TMS positioning, resulting in the intended stimulation intensities at the targeted cortical level.
format Online
Article
Text
id pubmed-10026380
institution National Center for Biotechnology Information
language English
publishDate 2022
record_format MEDLINE/PubMed
spelling pubmed-100263802023-03-20 Neuronavigation maximizes accuracy and precision in TMS positioning: Evidence from 11,230 distance, angle, and electric field modeling measurements Caulfield, Kevin A. Fleischmann, Holly H. Cox, Claire E. Wolf, Julia P. George, Mark S. McTeague, Lisa M. Brain Stimul Article BACKGROUND: Researchers and clinicians have traditionally relied on elastic caps with markings to reposition the transcranial magnetic stimulation (TMS) coil between trains and sessions. Newer neuronavigation technology co-registers the patient’s head and structural magnetic resonance imaging (MRI) scan, providing the researcher with real-time feedback about how to adjust the coil to be on-target. However, there has been no head to head comparison of accuracy and precision across treatment sessions. OBJECTIVE: /Hypothesis: In this two-part study, we compared elastic cap and neuronavigation targeting methodologies on distance, angle, and electric field (E-field) magnitude values. METHODS: In 42 participants receiving up to 50 total accelerated rTMS sessions in 5 days, we compared cap and neuronavigation targeting approaches in 3408 distance and 6816 angle measurements. In Experiment 1, TMS administrators saved an on-target neuronavigation location at Beam F3, which served as the landmark for all other measurements. Next, the operators placed the TMS coil based on cap markings or neuronavigation software to measure the distance and angle differences from the on-target sample. In Experiment 2, we saved each XYZ coordinate of the TMS coil from cap and neuronavigation targeting in 12 participants to compare the E-field magnitude differences at the cortical prefrontal target in 1106 cap and neuronavigation models. RESULTS: Cap targeting was significantly off-target for distance, placing the coil an average of 10.66 mm off-target (Standard error of the mean; SEM = 0.19 mm) compared to 0.3 mm (SEM = 0.03 mm) for neuronavigation (p < 0.0001). Cap targeting also significantly deviated for angles off-target, averaging 7.79 roll/pitch degrees (SEM = 1.07°) off-target and 5.99 yaw degrees (SEM = 0.12°) off-target; in comparison, neuronavigation targeting positioned the coil 0.34 roll/pitch degrees (SEM = 0.01°) and 0.22 yaw (SEM = 0.004°) off-target (both p < 0.0001). Further analyses revealed that there were significant inter-operator differences on distance and angle positioning for F3 (all p < 0.05), but not neuronavigation. Lastly, cap targeting resulted in significantly lower E-fields at the intended prefrontal cortical target, with equivalent E-fields as 110.7% motor threshold (MT; range = 58.3–127.4%) stimulation vs. 119.9% MT (range = 115–123.3%) from neuronavigated targeting with 120% MT stimulation applied (p < 0.001). CONCLUSIONS: Cap-based targeting is an inherent source of target variability compared to neuronavigation. Additionally, cap-based coil placement is more prone to differences across operators. Off-target coil placement secondary to cap-based measurements results in significantly lower amounts of stimulation reaching the cortical target, with some individuals receiving only 48.6% of the intended on-target E-field. Neuronavigation technology enables more precise and accurate TMS positioning, resulting in the intended stimulation intensities at the targeted cortical level. 2022 2022-08-27 /pmc/articles/PMC10026380/ /pubmed/36031059 http://dx.doi.org/10.1016/j.brs.2022.08.013 Text en https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) ).
spellingShingle Article
Caulfield, Kevin A.
Fleischmann, Holly H.
Cox, Claire E.
Wolf, Julia P.
George, Mark S.
McTeague, Lisa M.
Neuronavigation maximizes accuracy and precision in TMS positioning: Evidence from 11,230 distance, angle, and electric field modeling measurements
title Neuronavigation maximizes accuracy and precision in TMS positioning: Evidence from 11,230 distance, angle, and electric field modeling measurements
title_full Neuronavigation maximizes accuracy and precision in TMS positioning: Evidence from 11,230 distance, angle, and electric field modeling measurements
title_fullStr Neuronavigation maximizes accuracy and precision in TMS positioning: Evidence from 11,230 distance, angle, and electric field modeling measurements
title_full_unstemmed Neuronavigation maximizes accuracy and precision in TMS positioning: Evidence from 11,230 distance, angle, and electric field modeling measurements
title_short Neuronavigation maximizes accuracy and precision in TMS positioning: Evidence from 11,230 distance, angle, and electric field modeling measurements
title_sort neuronavigation maximizes accuracy and precision in tms positioning: evidence from 11,230 distance, angle, and electric field modeling measurements
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10026380/
https://www.ncbi.nlm.nih.gov/pubmed/36031059
http://dx.doi.org/10.1016/j.brs.2022.08.013
work_keys_str_mv AT caulfieldkevina neuronavigationmaximizesaccuracyandprecisionintmspositioningevidencefrom11230distanceangleandelectricfieldmodelingmeasurements
AT fleischmannhollyh neuronavigationmaximizesaccuracyandprecisionintmspositioningevidencefrom11230distanceangleandelectricfieldmodelingmeasurements
AT coxclairee neuronavigationmaximizesaccuracyandprecisionintmspositioningevidencefrom11230distanceangleandelectricfieldmodelingmeasurements
AT wolfjuliap neuronavigationmaximizesaccuracyandprecisionintmspositioningevidencefrom11230distanceangleandelectricfieldmodelingmeasurements
AT georgemarks neuronavigationmaximizesaccuracyandprecisionintmspositioningevidencefrom11230distanceangleandelectricfieldmodelingmeasurements
AT mcteaguelisam neuronavigationmaximizesaccuracyandprecisionintmspositioningevidencefrom11230distanceangleandelectricfieldmodelingmeasurements