Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
PURPOSE: The purpose of this study was to compare ultrasound pachymetry Ultrasound pachymetry (UP), specular microscopy (SM), and anterior segment optical coherence tomography (ASOCT) in the measurement of central corneal thickness (CCT) in cases of corneal edema before and after uncomplicated phaco...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10032283/ https://www.ncbi.nlm.nih.gov/pubmed/36968778 http://dx.doi.org/10.4103/sjopt.sjopt_63_20 |
_version_ | 1784910765609713664 |
---|---|
author | Artuç, Tuncay Batur, Muhammed |
author_facet | Artuç, Tuncay Batur, Muhammed |
author_sort | Artuç, Tuncay |
collection | PubMed |
description | PURPOSE: The purpose of this study was to compare ultrasound pachymetry Ultrasound pachymetry (UP), specular microscopy (SM), and anterior segment optical coherence tomography (ASOCT) in the measurement of central corneal thickness (CCT) in cases of corneal edema before and after uncomplicated phacoemulsification. METHODS: Preoperative pachymetric measurements were performed in all patients before cataract surgery. Sixty -seven patients who developed corneal edema after routine faoemulsification surgery were included in the study. CCT measurements were made the day before surgery and on the 1(st), 7(th), and 14(th) days after surgery. Correlation and agreement between the devices were quantified with the intraclass correlation coefficient (ICC), limits of agreement, and Bland–Altman plots. RESULTS: The mean age of the patients was 70.99 ± 9.56 (36–92) years. While CCT could be measured in all eyes with ASOCT after surgery, the number of eyes measured with UP and SM were 24 (11.7%) and 27 (16.1%), 55 (82.1%) and 52 (77.6%), and 46 (88.8%) and 45 (86.9%) on postoperative days 1, 7, and 14, respectively. ICC values between UP and SM, UP and ASOCT, and SM and ASOCT, respectively, were: preoperative 0.91, 0.94, and 0.92; 1(st) postoperative day 0.93, 0.86, and 0.93; 7(th) postoperative day 0.94, 0.85, and 0.96; and 14(th) postoperative day 0.95, 0.92, and 0.96. CONCLUSION: In all eyes with corneal edema, CCT measurements can be taken with ASOCT; however, in some cases, CCT measurements with SM and UP cannot be taken. However, there is a good correlation between measurements taken with these devices. |
format | Online Article Text |
id | pubmed-10032283 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-100322832023-03-23 Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography Artuç, Tuncay Batur, Muhammed Saudi J Ophthalmol Original Article PURPOSE: The purpose of this study was to compare ultrasound pachymetry Ultrasound pachymetry (UP), specular microscopy (SM), and anterior segment optical coherence tomography (ASOCT) in the measurement of central corneal thickness (CCT) in cases of corneal edema before and after uncomplicated phacoemulsification. METHODS: Preoperative pachymetric measurements were performed in all patients before cataract surgery. Sixty -seven patients who developed corneal edema after routine faoemulsification surgery were included in the study. CCT measurements were made the day before surgery and on the 1(st), 7(th), and 14(th) days after surgery. Correlation and agreement between the devices were quantified with the intraclass correlation coefficient (ICC), limits of agreement, and Bland–Altman plots. RESULTS: The mean age of the patients was 70.99 ± 9.56 (36–92) years. While CCT could be measured in all eyes with ASOCT after surgery, the number of eyes measured with UP and SM were 24 (11.7%) and 27 (16.1%), 55 (82.1%) and 52 (77.6%), and 46 (88.8%) and 45 (86.9%) on postoperative days 1, 7, and 14, respectively. ICC values between UP and SM, UP and ASOCT, and SM and ASOCT, respectively, were: preoperative 0.91, 0.94, and 0.92; 1(st) postoperative day 0.93, 0.86, and 0.93; 7(th) postoperative day 0.94, 0.85, and 0.96; and 14(th) postoperative day 0.95, 0.92, and 0.96. CONCLUSION: In all eyes with corneal edema, CCT measurements can be taken with ASOCT; however, in some cases, CCT measurements with SM and UP cannot be taken. However, there is a good correlation between measurements taken with these devices. Wolters Kluwer - Medknow 2023-03-09 /pmc/articles/PMC10032283/ /pubmed/36968778 http://dx.doi.org/10.4103/sjopt.sjopt_63_20 Text en Copyright: © 2023 Saudi Journal of Ophthalmology https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Artuç, Tuncay Batur, Muhammed Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography |
title | Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography |
title_full | Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography |
title_fullStr | Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography |
title_full_unstemmed | Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography |
title_short | Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography |
title_sort | comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10032283/ https://www.ncbi.nlm.nih.gov/pubmed/36968778 http://dx.doi.org/10.4103/sjopt.sjopt_63_20 |
work_keys_str_mv | AT artuctuncay comparisonofcentralcornealthicknessincornealedemabyultrasoundpachymetryspecularmicroscopyandanteriorsegmentopticalcoherencetomography AT baturmuhammed comparisonofcentralcornealthicknessincornealedemabyultrasoundpachymetryspecularmicroscopyandanteriorsegmentopticalcoherencetomography |