Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography

PURPOSE: The purpose of this study was to compare ultrasound pachymetry Ultrasound pachymetry (UP), specular microscopy (SM), and anterior segment optical coherence tomography (ASOCT) in the measurement of central corneal thickness (CCT) in cases of corneal edema before and after uncomplicated phaco...

Descripción completa

Detalles Bibliográficos
Autores principales: Artuç, Tuncay, Batur, Muhammed
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10032283/
https://www.ncbi.nlm.nih.gov/pubmed/36968778
http://dx.doi.org/10.4103/sjopt.sjopt_63_20
_version_ 1784910765609713664
author Artuç, Tuncay
Batur, Muhammed
author_facet Artuç, Tuncay
Batur, Muhammed
author_sort Artuç, Tuncay
collection PubMed
description PURPOSE: The purpose of this study was to compare ultrasound pachymetry Ultrasound pachymetry (UP), specular microscopy (SM), and anterior segment optical coherence tomography (ASOCT) in the measurement of central corneal thickness (CCT) in cases of corneal edema before and after uncomplicated phacoemulsification. METHODS: Preoperative pachymetric measurements were performed in all patients before cataract surgery. Sixty -seven patients who developed corneal edema after routine faoemulsification surgery were included in the study. CCT measurements were made the day before surgery and on the 1(st), 7(th), and 14(th) days after surgery. Correlation and agreement between the devices were quantified with the intraclass correlation coefficient (ICC), limits of agreement, and Bland–Altman plots. RESULTS: The mean age of the patients was 70.99 ± 9.56 (36–92) years. While CCT could be measured in all eyes with ASOCT after surgery, the number of eyes measured with UP and SM were 24 (11.7%) and 27 (16.1%), 55 (82.1%) and 52 (77.6%), and 46 (88.8%) and 45 (86.9%) on postoperative days 1, 7, and 14, respectively. ICC values between UP and SM, UP and ASOCT, and SM and ASOCT, respectively, were: preoperative 0.91, 0.94, and 0.92; 1(st) postoperative day 0.93, 0.86, and 0.93; 7(th) postoperative day 0.94, 0.85, and 0.96; and 14(th) postoperative day 0.95, 0.92, and 0.96. CONCLUSION: In all eyes with corneal edema, CCT measurements can be taken with ASOCT; however, in some cases, CCT measurements with SM and UP cannot be taken. However, there is a good correlation between measurements taken with these devices.
format Online
Article
Text
id pubmed-10032283
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-100322832023-03-23 Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography Artuç, Tuncay Batur, Muhammed Saudi J Ophthalmol Original Article PURPOSE: The purpose of this study was to compare ultrasound pachymetry Ultrasound pachymetry (UP), specular microscopy (SM), and anterior segment optical coherence tomography (ASOCT) in the measurement of central corneal thickness (CCT) in cases of corneal edema before and after uncomplicated phacoemulsification. METHODS: Preoperative pachymetric measurements were performed in all patients before cataract surgery. Sixty -seven patients who developed corneal edema after routine faoemulsification surgery were included in the study. CCT measurements were made the day before surgery and on the 1(st), 7(th), and 14(th) days after surgery. Correlation and agreement between the devices were quantified with the intraclass correlation coefficient (ICC), limits of agreement, and Bland–Altman plots. RESULTS: The mean age of the patients was 70.99 ± 9.56 (36–92) years. While CCT could be measured in all eyes with ASOCT after surgery, the number of eyes measured with UP and SM were 24 (11.7%) and 27 (16.1%), 55 (82.1%) and 52 (77.6%), and 46 (88.8%) and 45 (86.9%) on postoperative days 1, 7, and 14, respectively. ICC values between UP and SM, UP and ASOCT, and SM and ASOCT, respectively, were: preoperative 0.91, 0.94, and 0.92; 1(st) postoperative day 0.93, 0.86, and 0.93; 7(th) postoperative day 0.94, 0.85, and 0.96; and 14(th) postoperative day 0.95, 0.92, and 0.96. CONCLUSION: In all eyes with corneal edema, CCT measurements can be taken with ASOCT; however, in some cases, CCT measurements with SM and UP cannot be taken. However, there is a good correlation between measurements taken with these devices. Wolters Kluwer - Medknow 2023-03-09 /pmc/articles/PMC10032283/ /pubmed/36968778 http://dx.doi.org/10.4103/sjopt.sjopt_63_20 Text en Copyright: © 2023 Saudi Journal of Ophthalmology https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Artuç, Tuncay
Batur, Muhammed
Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
title Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
title_full Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
title_fullStr Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
title_full_unstemmed Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
title_short Comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
title_sort comparison of central corneal thickness in corneal edema by ultrasound pachymetry, specular microscopy, and anterior segment optical coherence tomography
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10032283/
https://www.ncbi.nlm.nih.gov/pubmed/36968778
http://dx.doi.org/10.4103/sjopt.sjopt_63_20
work_keys_str_mv AT artuctuncay comparisonofcentralcornealthicknessincornealedemabyultrasoundpachymetryspecularmicroscopyandanteriorsegmentopticalcoherencetomography
AT baturmuhammed comparisonofcentralcornealthicknessincornealedemabyultrasoundpachymetryspecularmicroscopyandanteriorsegmentopticalcoherencetomography