Comparison of central corneal thickness measurements obtained by OrbscanIIz(®) and ultrasound pachymetry: A concordance study in the usual clinical practice
PURPOSE: The objective of this study was to assess the concordance between the values obtained in measuring central corneal thickness using the OrbscanIIz(®) and the contact ultrasonic pachymeter available in our public ophthalmology service. METHODS: Measurements were taken from 88 eyes of 44 patie...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10032292/ https://www.ncbi.nlm.nih.gov/pubmed/36968780 http://dx.doi.org/10.4103/sjopt.Sjopt_82_21 |
Sumario: | PURPOSE: The objective of this study was to assess the concordance between the values obtained in measuring central corneal thickness using the OrbscanIIz(®) and the contact ultrasonic pachymeter available in our public ophthalmology service. METHODS: Measurements were taken from 88 eyes of 44 patients using the two instruments. The data obtained were statistically analyzed using version 22 of the IBM SPSS(®) program. RESULTS: The mean of central corneal thickness measurements obtained from OrbscanIIz(®) was significantly higher than that obtained from ultrasound pachymetry. However, the mean of differences between both instruments was only 7.22 μ, which could be considered a clinically insignificant result when considering the good concordance obtained between both systems. CONCLUSION: OrbscanIIz(®) and ultrasound pachymetry can be interchangeable in the usual public clinical practice when measuring central corneal thickness. This is the first research found in the literature that uses a concordance study to compare the data resulting from central corneal thickness measurements obtained by an OrbscanIIz(®) and an OcuScan(®) pachymeter in our environment. |
---|